First clinical evaluation of a new percutaneous optical fiber glucose sensor for continuous glucose monitoring in diabetes.
نویسندگان
چکیده
BACKGROUND This article describes a new fiber-coupled, percutaneous fluorescent continuous glucose monitoring (CGM) system that has shown 14 days of functionality in a human clinical trial. METHOD The new optical CGM system (FiberSense) consists of a transdermal polymer optical fiber containing a biochemical glucose sensor and a small fluorescence photometer optically coupled to the fiber. The glucose-sensitive optical fiber was implanted in abdominal and upper-arm subcutaneous tissue of six diabetes patients and remained there for up to 14 days. The performance of the system was monitored during six visits to the study center during the trial. Blood glucose changes were induced by oral carbohydrate intake and insulin injections, and capillary blood glucose samples were obtained from the finger tip. The data were analyzed using linear regression and the consensus error grid analysis. RESULTS The FiberSense worn at the upper arm exhibited excellent results during 14 wearing days, with an overall mean absolute relative difference (MARD) of 8.3% and 94.6% of the data in zone A of the consensus error grid. At the abdominal application site, FiberSense resulted in a MARD of 11.4 %, with 93.8% of the data in zone A. CONCLUSIONS The FiberSense CGM system provided consistent, reliable measurements of subcutaneous glucose levels in human clinical trial patients with diabetes for up to 14 days.
منابع مشابه
Determination of Blood Glucose Concentration by Using Wavelet Transform and Neural Networks
Background: Early and non-invasive determination of blood glucose level is of great importance. We aimed to present a new technique to accurately infer the blood glucose concentration in peripheral blood flow using non-invasive optical monitoring system. Methods: The data for the research were obtained from 900 individuals. Of them, 750 people had diabetes mellitus (DM). The system was ...
متن کاملThirty-fifth anniversary of the optical affinity sensor for glucose: a personal retrospective.
Since 1962 when Clark introduced the enzyme electrode, research has been intense for a robust implantable glucose sensor. An alternative "optical affinity sensor" was introduced by Jerome Schultz in 1979. The evolution of this sensor technology into a new methodology is reviewed. The approach integrates a variety of disparate concepts: the selectivity of immunoassays-selectivity for glucose was...
متن کاملNon-invasive Optical Techniques for determination of blood Glucose levels: A Review Article
This article reviews the development of non-invasive optical techniques for determination of blood glucose concentrations in diabetic patients. Early diagnosis and daily management are essential for ensuring the healthy life of diabetic patients. The determination of blood glucose concentration with common devices involves the chemical analysis of blood samples, which are obtained by pricking t...
متن کاملPercutaneous fiber-optic sensor for chronic glucose monitoring in vivo.
We are developing a family of fiber-optic sensors called Sencils (sensory cilia), which are disposable, minimally invasive, and can provide in vivo monitoring of various analytes for several weeks. The key element is a percutaneous optical fiber that permits reliable spectroscopic measurement of chemical reactions in a nano-engineered polymeric matrix attached to the implanted end of the fiber....
متن کاملA fiber optic sensor for measuring glucose in aqueous solutions
In this paper, the set-up of a multi-mode fiber optic sensor for measuring glucose in aqueous solutions is investigated and evaluated. The basis of this sensor is based on the Fresnel Reflection. In this setup, a helium-neon laser is used as a light source, a fiber optic probe, photocell as detector and a digital multimeter. The statistical analysis of the recorded data shows a highly linear be...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of diabetes science and technology
دوره 7 1 شماره
صفحات -
تاریخ انتشار 2013