Performance characteristics of loci method for measuring cardiac troponin I on the dimension EXL

نویسندگان

  • José L. Martín Calderón
  • Julia M. Varona Pérez
  • Fernando Bustos
  • Juan C. Sánchez Gómez
چکیده

OBJECTIVE To define the performance characteristics of the LOCI® method for cardiac troponin I on the Dimension EXL system. DESIGNS AND METHODS Three different levels of commercial control (mean concentrations 0.426, 1.42, and 18.64 µg/L) were used for the imprecision study, quantifying separately within-run and between-run over 20 days. The limit of blank (LoB) and limit of detection (LoD) were assessed with 20 replicates of a sample without troponin I. Linearity was assessed by regression analysis. In addition, we studied inaccuracy, carry-over and limit of quantitation and conducted a method comparison with the Stratus CS (n=69). The reference interval was determined in 146 healthy blood donors using non-parametric method. RESULTS The within-run imprecision (coefficient of variation [CV], %) obtained at each level was 2.4, 1.4% and 2.2%, while the between-run imprecision (CV,%) was 3.3%, 2.9% and 2.5%. Total imprecision was 4.06%, 3.3% and 3.4% for each control level. The limit of quantitation which corresponds to the troponin I concentration at which CV=10% was 0.05 µg/L. Method comparison with the Stratus CS assay produced the equation: Dimension EXL=-0.002698+1.0233⁎(Stratus CS) with a confidence interval from -0.01562 to 0.00626 for the intercept and (0.979 to 1.0875) for the slope. The 99th percentile obtained for the reference population was 0.047 µg/L. CONCLUSIONS The LOCI method for cardiac troponin I on the Dimension EXL meets all guidelines recommended criteria referring to limit of quantitation, imprecision and shows excellent transferability with the Stratus CS method.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An Aptamer-based Biosensor for Troponin I Detection in Diagnosis of Myocardial Infarction

Background: Acute myocardial infarction (MI) accounts for one third of deaths. Cardiac troponin I (TnI) is a reliable biomarker of cardiac muscle tissue injury and is employed in the early diagnosis of MI.Objectives: In this study, a molecular method is introduced to early diagnosis of MI by rapid detection of TnI.Materials and Methods: The detection method was based on electrochemical aptasens...

متن کامل

The most sensitive double-marker method of cardiac isoenzymes of creatinine kinase, troponin I, and myoglobin in different time periods for diagnosing acute myocardial infarction

  Purpose: Myocardial infarction is among the most common diagnoses in patients admitted   to hospitals in western countries, and its rapid diagnosis is of utmost importance. This study   was conducted to determine the most sensitive double-marker cardiac isoenzyme of creatinine   kinase, troponin I, and myoglobin for diagnosing acute myocardial infarction.   Materials and Methods: This was an ...

متن کامل

Performance of a revised cardiac troponin method that minimizes interferences from heterophilic antibodies.

BACKGROUND Recent guidelines for use of cardiac troponin to detect cardiac damage and for cardiovascular risk stratification have made increasingly sensitive troponin assays important. Troponin assays continue to be plagued by interferences caused by heterophilic antibodies (HAs). We evaluated the performance of a revised cardiac troponin I (cTnI) assay designed to have increased analytical sen...

متن کامل

Analysis of the relation between coping ways with stress and Cardiac Biomarker Troponin I in coronary heart disease patients

Introduction: Coping ways with stress in coronary heart disease patients can lead to significant changes in levels of Cardiac Biomarker Troponin I, but researchers had not explored it empirically. So, the main objective of this study was to identify the relationship between coping ways with stress and Cardiac Biomarker Troponin I and also, finding coping ways that predict changes in the concent...

متن کامل

Values of troponin T and myoglobin predictive of non-cardiac ischemia in rats

Objective(s):Biochemical markers are important for the timely diagnosis and follow-up of ischemic events. Most of the markers have been previously studied in the context of cardiac ischemia. However, research on markers of non-cardiac events has been insufficient. Therefore, we investigated the relationship between troponin and myoglobin which are commonly used markers of cardiac ischemia, in n...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 1  شماره 

صفحات  -

تاریخ انتشار 2015