Comparison of finger and intra-arterial blood pressure monitoring at rest and during laboratory testing.
نویسندگان
چکیده
The accuracy of blood pressure values obtained by continuous noninvasive finger blood pressure recording via the FINAPRES device was evaluated by comparison with simultaneous intraarterial monitoring both at rest and during performance of tests known to induce fast and often marked changes in blood pressure. The comparison was performed in 24 normotensive or essential hypertensive subjects. The average discrepancy between finger and intra-arterial blood pressure recorded over a 30-minute rest period was 6.5 +/- 2.6 mm Hg and 5.4 +/- 2.9 mm Hg for systolic and diastolic blood pressure, respectively; a close between-method correspondence was also demonstrated by linear regression analysis. The beat-to-beat changes in finger systolic and diastolic blood pressure were on average similar to those measured intra-arterially during tests that induced a pressor or depressor response (hand-grip, cold pressor test, diving test, Valsalva maneuver, intravenous injections of phenylephrine and trinitroglycerine) as well as during tests that caused vasomotor changes without major variations in blood pressure (application of lower body negative pressure, passive leg raising). The average between-method discrepancy in the evaluation of blood pressure changes was never greater than 4.3 and 2.0 mm Hg for systolic and diastolic blood pressure, respectively; the corresponding standard deviations ranged between 4.6 and 1.6 mm Hg. Beat-to-beat computer analysis of blood pressure variability over the 30-minute rest period provided standard deviations almost identical when calculated by separate consideration of intra-arterial and finger blood pressure tracings (3.7 and 3.8 mm Hg, respectively).(ABSTRACT TRUNCATED AT 250 WORDS)
منابع مشابه
Comparison of direct and indirect blood pressure in patients under esophagectomy surgery: a prospective observational study
Background: Arterial pressure is one of the most important physiological variables and often needs to be monitored repeatedly or continuously in perioperative period. Arterial pressure monitoring is one of the standard monitoring in operating room. During general anesthesia, blood pressure can be measured by using a noninvasive arterial pressure method or continuous invasive arterial pressure b...
متن کاملSpectral and sequence analysis of finger blood pressure variability. Comparison with analysis of intra-arterial recordings.
The aim of our study was to assess whether the Finapres device is able to accurately monitor not only average blood pressure values but also blood pressure variability. To examine this issue, we analyzed 30-minute recordings of finger and intra-arterial pressure simultaneously obtained at rest in 14 patients. We compared systolic blood pressure, diastolic blood pressure, mean arterial pressure,...
متن کاملEvaluation of noninvasive blood pressure monitoring devices Spacelabs 90202 and 90207 versus resting and ambulatory 24-hour intra-arterial blood pressure.
This study evaluated the accuracy of blood pressure values provided by the Spacelabs 90202 and 90207 devices in comparison with intra-arterial recording in 19 subjects at rest and in nine subjects in ambulatory conditions (Oxford method). At rest Spacelabs monitors reflected intra-arterial systolic blood pressure values very closely but overestimated to a considerable extent intra-arterial dias...
متن کاملA study of variation in cardiocirculatory parameters with different body positions during isometric exercise in young adult males.
Introduction: To study the effects of exercise, how important it is to choose a posture? We aimed to characterize the possible impact of different body positions on cardiovascular parameters during and after sustained isometric handgrip (IHG) exercise. Materials and Methods: Cross sectional study was carried out in 33 young adult males (mean age: 19.21±1.083 years). We recorded Bloo...
متن کاملSynthesis of Novel 4-[1-(4-fluorobenzyl)-5-imidazolyl] Dihydropyridines and Studying their Effects on Rat Blood Pressure
Objective(s) Calcium-channel blockers have an important role in the treatment of several cardiovascular disorders. Derivatives of 1, 4-dihydropyridine are one of the most potent calcium antagonists. In this study four novel ', 4-dihydropyridine calcium channel blockers were synthesized and their hypotensive properties were investigated in male rats. Materials and Methods Four 1, 4-dihydropyr...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Hypertension
دوره 13 6 Pt 1 شماره
صفحات -
تاریخ انتشار 1989