Increased risk of revision of cementless stemmed total hip arthroplasty with metal-on-metal bearings
نویسندگان
چکیده
BACKGROUND AND PURPOSE Data from the national joint registries in Australia and England and Wales have revealed inferior medium-term survivorship for metal-on-metal (MoM) total hip arthroplasty (THA) than for metal-on-polyethylene (MoP) THA. Based on data from the Nordic Arthroplasty Register Association (NARA), we compared the revision risk of cementless stemmed THA with MoM and MoP bearings and we also compared MoM THA to each other. PATIENTS AND METHODS We identified 32,678 patients who were operated from 2002 through 2010 with cementless stemmed THA with either MoM bearings (11,567 patients, 35%) or MoP bearings (21,111 patients, 65%). The patients were followed until revision, death, emigration, or the end of the study period (December 31, 2011), and median follow-up was 3.6 (interquartile range (IQR): 2.4-4.8) years for MoM bearings and 3.4 (IQR: 2.0-5.8) years for MoP bearings. Multivariable regression in the presence of competing risk of death was used to assess the relative risk (RR) of revision for any reason (with 95% confidence interval (CI)). RESULTS The cumulative incidence of revision at 8 years of follow-up was 7.0% (CI: 6.0-8.1) for MoM bearings and 5.1% (CI: 4.7-5.6) for MoP bearings. At 6 years of follow-up, the RR of revision for any reason was 1.5 (CI: 1.3-1.7) for MoM bearings compared to MoP bearings. The RR of revision for any reason was higher for the ASR (adjusted RR = 6.4, CI: 5.0-8.1), the Conserve Plus (adjusted RR = 1.7, CI: 1.1-2.5) and "other" acetabular components (adjusted RR = 2.4, CI: 1.5-3.9) than for MoP THA at 6 years of follow-up. INTERPRETATION At medium-term follow-up, the survivorship for cementless stemmed MoM THA was inferior to that for MoP THA, and metal-related problems may cause higher revision rates for MoM bearings with longer follow-up.
منابع مشابه
Comparison of the risk of revision in cementless total hip arthroplasty with ceramic-on-ceramic and metal-on-polyethylene bearings
BACKGROUND AND PURPOSE Ceramic-on-ceramic (CoC) bearings were introduced in total hip arthroplasty (THA) to reduce problems related to polyethylene wear. We compared the 9-year revision risk for cementless CoC THA and for cementless metal-on-polyethylene (MoP) THA. PATIENTS AND METHODS In this prospective, population-based study from the Danish Hip Arthroplasty Registry, we identified all the...
متن کاملTotal Hip Replacement Revision in a Single Brand Small Cementless Stem – Our Experience after the Findings of the National Joint Registry
Background: Cementless total hip replacement is the common THR performed in England, Wales, Northern Irelandand the Isle of Man. The Corail stem is the most popular cementless implant and has a ODEP 10A rating. Review ofits performance in the registry identified an increase rate of revision amongst the smaller stem sizes. However, claritywas not provided on the explanation for this finding. We ...
متن کاملCementless Total Hip Arthroplasty Involving Trochanteric Osteotomy without Subtrochanteric Shortening for High Hip Dislocation
BACKGROUND Total hip arthroplasty with subtrochanteric shortening osteotomy is widely performed for high hip dislocation. However, suboptimal leg length discrepancy correction and nonunion of the osteotomy site remain concerns. Although total hip arthroplasty using trochanteric osteotomy without subtrochanteric osteotomy was introduced, cemented implants have been more commonly used than contem...
متن کاملLong-Term Results of Cementless Total Hip Arthroplasty Using a Modular Stem in Patients with Femoral Head Osteonecrosis
Cementless modular stemwas designed to allow the surgeon to independently match the anatomy of the metaphyseal and diaphyseal regions of the femoral canal of an individual patient.1 S-ROM stem (DePuy/Johnson & Johnson), which has a hollow proximal sleeve and separate cylindrical core, was introduced in 1984. So far, long-term results of this prosthesis have been excellent in osteoarthritis, dev...
متن کاملDistributed analysis of hip implants using six national and regional registries: comparing metal-on-metal with metal-on-highly cross-linked polyethylene bearings in cementless total hip arthroplasty in young patients.
BACKGROUND The regulation of medical devices has attracted controversy recently because of problems related to metal-on-metal hip implants. There is growing evidence that metal-on-metal implants fail early and cause local and systemic complications. However, the failure associated with metal-on-metal head size is not consistently documented and needs to be communicated to patients and surgeons....
متن کامل