Common hemoglobin variants in southern Taiwan and their effect on the determination of HbA1c by ion-exchange high-performance liquid chromatography.

نویسندگان

  • Chih-Hsun Chu
  • Hing-Chung Lam
  • Jenn-Kuen Lee
  • Mei-Chun Wang
  • Chih-Chen Lu
  • Chun-Chin Sun
  • Ming-Ju Chuang
چکیده

BACKGROUND Patients with hemoglobin (Hb) variants may produce false HbA1c measurement. This study aimed to detect the common Hb variants in southern Taiwan and to evaluate their effect on the determination of HbA1c. METHODS A total of 1,434 samples collected for HbA1c measurement at Kaohsiung Veterans General Hospital in southern Taiwan in March 2008 were submitted for Hb variant analysis by Primus CLC-385. HbA1c measurements were obtained using ion-exchange high-performance liquid chromatography (HPLC) (Tosoh HLC-723 G7) for routine analysis. Patients identified with Hb variants were recalled for boronate-affinity HPLC analysis. The values of estimated average glucose (eAG) were converted from HbA1c. Values of eAG-FPG, calculated by eAG minus fasting plasma glucose (FPG), were compared to estimate the accuracy of HbA1c measurement in patients with Hb variants. RESULTS Among the 1,434 patients, the mean standard deviation of FPG was 162.8 +/- 60.5 mg/dL, HbA1c was 8.28 +/- 1.97%, and eAG was 190.9 +/- 56.6 mg/dL. Five Hb variants were detected in 11 patients, the incidence being 0.76%. Hb J was identified in 4 patients, Hb G in 2 patients, Hb E in 1 patient, Hb owari in 3 patients, and high fetal hemoglobin (HbF) in 1 patient. Abnormal HPLC chromatograms were seen among the patients with Hb J, E, G and HbF, but not in the patients with Hb owari. In patients with Hb variants, FPG was 149.5 +/- 39.9 mg/dL, HbA1c was 7.29 +/- 2.01%, and eAG was 162.5 +/- 57.7 mg/dL. Lower values of eAG-FPG may have occurred in the patients with Hb J and E, and in those with high HbF. On scattergrams of the relationship between HbA1c and FPG, the plots of Hb J, E and high HbF lay below the regression line of non-Hb variants. Inconsistent Hb values between both methods were only observed among some samples of patients with Hb variants. CONCLUSION The existence of Hb variants may result in false HbA1c measurement. The possible presence of spuriously low HbA1c levels or abnormal HPLC chromatograms by using ion-exchange methods should be kept in mind.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A review of variant hemoglobins interfering with hemoglobin A1c measurement.

Hemoglobin A1c (HbA1c) is used routinely to monitor long-term glycemic control in people with diabetes mellitus, as HbA1c is related directly to risks for diabetic complications. The accuracy of HbA1c methods can be affected adversely by the presence of hemoglobin (Hb) variants or elevated levels of fetal hemoglobin (HbF). The effect of each variant or elevated HbF must be examined with each sp...

متن کامل

Management of Hemoglobin Variants Detected Incidentally in HbA1c Testing: A Common Problem Currently Lacking a Standard Approach

Although HbA1c is widely used for diagnosis and management of diabetes in clinical practice and research studies, management of the incidental detection of hemoglobin variants (e.g., hemoglobin S [HbS], HbC) by common HbA1c methods, such as ion exchange high-performance liquid chromatography (HPLC), is not well established. Many clinical laboratories do not report the presence of hemoglobin var...

متن کامل

Ionic Liquid-based Ultrasound-assisted In-situ Solvent Formation Microextraction and High-performance Liquid Chromatography for the Trace Determination of Polycyclic Aromatic Hydrocarbons in Environmental Water Samples

A green and efficient ionic liquid-based ultrasound-assisted in-situ solvent formationmicroextraction (IL-UA-ISFME) in combination with high-performance liquid chromatographyultravioletdetection (HPLC-UV) has been successfully developed for the trace determination offive selected polycyclic aromatic hydrocarbons (PAHs) in environmental water samples. In thismethod, a hydrophobic ionic liquid (1...

متن کامل

Effect of HbE trait on measurement of HbA1c by three different methods.

Sir, HbA1c testing is mainly used for monitoring glycaemic control in patients with diabetes. However, the World Health Organization now recommends that HbA1c can be used as a diagnostic test for diabetes, provided that stringent quality assurance procedures are in place and assays are standardised to criteria aligned to the international reference values. Laboratories use many different method...

متن کامل

Hb H Interference on Measurement Of HbA1c With Ion-Exchange HPLC

In this article, an interference caused by hemoglobin H (Hb H), during the measurement of hemoglobin A1c (HbA1c) with ion exchange high pressure liquid chromatography (HPLC) method, was presented in blood sample of a 20-year-old male patient. HbA1c measurement was performed with Agilent 1200 HPLC system using a commercial Recipe HbA1c ion-exchange column. Hemoglobin electrophoresis was performe...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of the Chinese Medical Association : JCMA

دوره 72 7  شماره 

صفحات  -

تاریخ انتشار 2009