INTRODUCTION Information about implantable cardioverter-defibrillator (ICD) longevity is mostly calculated from measurements under ideal laboratory conditions. However, little information about longevity under clinical circumstances is available. This survey gives an overview on ICD service times and generator replacements in a cohort of consecutive ICD patients. METHODS Indications for replacement were classified as a normal end-of-service (EOS), premature EOS, system malfunction, infection and device advisory, or recall actions. From the premature and normal EOS group, longevity from single-chamber (SC), dual-chamber (DC), and cardiac resynchronization therapy defibrillator (CRT-D), rate-responsive (RR) settings, high output (HO) stimulation, and indication for ICD therapy was compared. Differences between brands were compared as well. RESULTS In a total of 854 patients, 203 ICD replacements (165 patients) were recorded. Premature and normal EOS replacements consisted of 32 SC, 98 DC and 24 CRT-D systems. Longevity was significantly longer in SC systems compared to DC and CRT-D systems (54 +/- 19 vs. 40 +/- 17 and 42 +/- 15 months; P = 0.008). Longevity between non-RR (n = 143) and RR (n = 11) settings was not significantly different (43 +/- 18 vs. 45 +/- 13 months) as it also was not for HO versus non-HO stimulation (43 +/- 19 vs. 46 +/- 17 months). Longevity of ICDs was not significantly different between primary and secondary prevention (42 +/- 19 vs. 44 +/- 18 months). The average longevity on account of a device-based EOS message was 43 +/- 18 months. Average longevity for Biotronik (BIO, n = 72) was 33 +/- 10 months, for ELA Medical (ELA, n = 12) 44 +/- 17 months, for Guidant (GDT, n = 36) 49 +/- 12 months, for Medtronic (MDT, n = 29) 62 +/- 22 months, and for St. Jude Medical (SJM, n = 5) 31 +/- 9 months (P < 0.001). CONCLUSION SC ICD generators had a longer service time compared to DC and CRT-D systems. No influence of indication for ICD therapy and HO stimulation on generator longevity was observed in this study. MDT ICDs had the longest service time.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Implantable cardioverter-defibrillator in a patient with dextrocardia situs inversus

Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...

متن کامل

زندگی با دفیبریلاتور قابل کاشت: یک مطالعه پدیده‌شناسی

  Background & Aim: Implantable Cardioverter Defibrillator is the most effective treatment for cardiac dysrhythmias. Having such life-saving device is the unique experience with important changes in the patient’ life. Understanding of these experiences can lead to success in treatment and follow- up.This study aimed to understand the experiences of patients with Implantable Cardioverter Defibri...

متن کامل

THE EFFECT OF NURSE-PEER-LED SUPPORT INTERVENTION ON QUALITY OF LIFE IN PATIENTS WITH AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR: A RANDOMIZED CLINICAL TRIAL

Background & Aims: Nowadays, the quality of life and how coping and treatment of patients with an implantable cardiac defibrillator has led to concern and attempts to reduce or eliminate the problems of these patients in the people who provide health care services. This study aimed to determine the effect of nurse-peer-led support intervention on quality of life in patients with an implantable ...

متن کامل

An introduction to the automatic implantable cardioverter-defibrillator and its anesthetic implications.

This paper presents a description of the automatic implantable cardioverter-defibrillator (AICD), as well as information pertaining to indications for its use, patient selection criteria, surgical implantation techniques, and anesthetic implications. A brief historical overview of the development of the AICD is included. Complications, indications, and contraindications associated with the devi...

متن کامل

Battery longevity in cardiac resynchronization therapy implantable cardioverter defibrillators.

AIMS Cardiac resynchronization therapy (CRT) implantable cardioverter defibrillators (ICDs) deliver high burden ventricular pacing to heart failure patients, which has a significant effect on battery longevity. The aim of this study was to investigate whether battery longevity is comparable for CRT-ICDs from different manufacturers in a contemporary cohort of patients. METHODS AND RESULTS All...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Pacing and clinical electrophysiology : PACE

دوره 32 10  شماره 

صفحات  -

تاریخ انتشار 2009