In-air calibration of an HDR 192Ir brachytherapy source using therapy ion chambers.

نویسندگان

  • Narayan Prasad Patel
  • Bishnu Majumdar
  • V Vijiyan
  • Pradeep K Hota
چکیده

The Gammamed Plus 192Ir high dose rate brachytherapy sources were calibrated using the therapy level ionization chambers (0.1 and 0.6 cc) and the well-type chamber. The aim of the present study was to assess the accuracy and suitability of use of the therapy level chambers for in-air calibration of brachytherapy sources in routine clinical practice. In a calibration procedure using therapy ion chambers, the air kerma was measured at several distances from the source in a specially designed jig. The room scatter correction factor was determined by superimposition method based on the inverse square law. Various other correction factors were applied on measured air kerma values at multiple distances and mean value was taken to determine the air kerma strength of the source. The results from four sources, the overall mean deviation between measured and quoted source strength by manufacturers was found -2.04% (N = 18) for well-type chamber. The mean deviation for the 0.6 cc chamber with buildup cap was found -1.48 % (N = 19) and without buildup cap was 0.11% (N = 22). The mean deviation for the 0.1 cc chamber was found -0.24% (N = 27). Result shows that probably the excess ionization in case of 0.6 cc therapy ion chamber without buildup cap was estimated about 2.74% and 1.99% at 10 and 20 cm from the source respectively. Scattered radiation measured by the 0.1 cc and 0.6 cc chamber at 10 cm measurement distance was about 1.1% and 0.33% of the primary radiation respectively. The study concludes that the results obtained with therapy level ionization chambers were extremely reproducible and in good agreement with the results of the well-type ionization chamber and source supplier quoted value. The calibration procedure with therapy ionization chambers is equally competent and suitable for routine calibration of the brachytherapy sources.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy

Background: The effectiveness and safety of brachytherapy treatment is mainly concerned with the calibration of sources and their traceability to internationally accepted Standards. Secondary Standard Dosimetry Laboratory (SSDL) does not offer calibration of ionization chambers with gamma-ray spectrum of high dose rate source 192Ir . This work has been carried out to calibrate the high dose rat...

متن کامل

A practical implementation of the 2010 IPEM high dose rate brachytherapy code of practice for the calibration of 192Ir sources.

This paper details a practical method for deriving the reference air kerma rate calibration coefficient for Farmer NE2571 chambers using the U.K. Institute of Physics and Engineering in Medicine (IPEM) code of practice for the determination of the reference air kerma rate for HDR (192)Ir brachytherapy sources based on the National Physical Laboratory (NPL) air kerma standard. The reference air ...

متن کامل

Dosimetric characterization of round HDR 192Ir accuboost applicators for breast brachytherapy.

PURPOSE The AccuBoost brachytherapy system applies HDR 192Ir beams peripherally to the breast using collimating applicators. The purpose of this study was to benchmark Monte Carlo simulations of the HDR 192Ir source, to dosimetrically characterize the round applicators using established Monte Carlo simulation and radiation measurement techniques and to gather data for clinical use. METHODS Do...

متن کامل

Verification of delivered dose to the Lips carcinoma tumors with HDR brachytherapy sources: 192Ir and 60CO in an exclusive plexiglass Phantom.

Introduction: Brachytherapy, especially using manually after loaded Iridium-192 and Cobalt-60, can be applied as a sole treatment, as a treatment complementary to surgery, and as a local boost in combination with EBRT. The use of HDR brachytherapy catheters incorporated in removable dental molds allows repeated, highly reproducible, fractionated outpatient brachytherapy of supe...

متن کامل

Comparison the treatment time of cobalt and iridium sources in HDR brachytherapy

Introduction: The use of long-half-life Cobalt source instead of iridium source will be very helpful in HDR brachytherapy for treatment of cervical cancer. The aim of this study was compared the treatment time with cobalt and iridium source in brachytherapy for treatment of cervical cancer using Tandem-ovoid applicator. Materials and Methods: In this study, 19...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of cancer research and therapeutics

دوره 1 4  شماره 

صفحات  -

تاریخ انتشار 2005