Improvement in PET/CT image quality with a combination of point-spread function and time-of-flight in relation to reconstruction parameters.

نویسندگان

  • Go Akamatsu
  • Kaori Ishikawa
  • Katsuhiko Mitsumoto
  • Takafumi Taniguchi
  • Nobuyoshi Ohya
  • Shingo Baba
  • Koichiro Abe
  • Masayuki Sasaki
چکیده

UNLABELLED The aim of this study was to investigate the effects of the point-spread function (PSF) and time-of-flight (TOF) on improving (18)F-FDG PET/CT images in relation to reconstruction parameters and noise-equivalent counts (NEC). METHODS This study consisted of a phantom study and a retrospective analysis of 39 consecutive patients who underwent clinical (18)F-FDG PET/CT. The body phantom of the National Electrical Manufacturers Association and International Electrotechnical Commission with a 10-mm-diameter sphere was filled with an (18)F-FDG solution with a 4:1 radioactivity ratio compared with the background. The PET data were reconstructed with the baseline ordered-subsets expectation maximization (OSEM) algorithm, with the OSEM+PSF model, with the OSEM+TOF model, and with the OSEM+PSF+TOF model. We evaluated image quality by visual assessment, the signal-to-noise ratio of the 10-mm sphere (SNR(10 mm)), the contrast of the 10-mm sphere, and the coefficient of variance in the phantom study and then determined the optimal reconstruction parameters. We also examined the effects of PSF and TOF on the quality of clinical images using the signal-to-noise ratio in the liver (SNR(liver)) in relation to the NEC in the liver (NEC(liver)). RESULTS In the phantom study, the SNR(10 mm) was the highest for the OSEM+PSF+TOF model, and the highest value was obtained at iteration 2 for algorithms with the TOF and at iteration 3 for those without the TOF. In terms of a postsmoothing filter full width at half maximum (FWHM), the high SNR(10 mm) was obtained with no filtering or was smaller than 2 mm for algorithms with PSF and was 4-6 mm for those without PSF. The balance between the contrast recovery and noise is different for algorithms with either PSF or TOF. A combination of PSF and TOF improved SNR(10 mm), contrast, and coefficient of variance, especially with a small-FWHM gaussian filter. In the clinical study, the SNR(liver) of the low-NEC(liver) group in the OSEM+PSF+TOF model was compared with that of the high-NEC(liver) group in conventional OSEM. The PSF+TOF improved the SNR(liver) by about 24.9% ± 9.81%. CONCLUSION A combination of PSF and TOF clearly improves image quality, whereas optimization of the reconstruction parameters is necessary to obtain the best performance for PSF or TOF. Furthermore, this combination has the potential to provide good image quality with either lower activity or shorter acquisition time, thus improving patient comfort and reducing the radiation burden.

منابع مشابه

Edge Artifacts in Point Spread Function-based PET Reconstruction in Relation to Object Size and Reconstruction Parameters

Objective(s): We evaluated edge artifacts in relation to phantom diameter and reconstruction parameters in point spread function (PSF)-based positron emission tomography (PET) image reconstruction.Methods: PET data were acquired from an original cone-shaped phantom filled with 18F solution (21.9 kBq/mL) for 10 min using a Biograph mCT scanner. The images were reconstructed using the baseline or...

متن کامل

Evaluation of the Effect of Tumor Position on Standardized Uptake Value Using Time-of-Flight Reconstruction and Point Spread Function

Objective(s): The present study was conducted to examine whether the standardized uptake value (SUV) may be affected by the spatial position of a lesion in the radial direction on positron emission tomography (PET) images, obtained via two methods based on time-of-flight (TOF) reconstruction and point spread function (PSF). Methods: A cylinder phantom with the sphere (30mm diameter), located in...

متن کامل

بررسی تاثیر کرنل های مختلف بازسازی تصویر بر روی نویز و رزولوشن فضایی تصاویر سی تی اسکن: مطالعه بر روی فانتوم

Introduction: Noise and spatial resolution (SR) are the main characteristics of image quality in CT scanning affected by different radiation and technical parameters. One of the important parameters are reconstruction kernels. Different reconstruction kernels are used for noise reduction and/or edge-enhancement purposes. In this study, we investigate some reconstruction kern...

متن کامل

Impact of Various Image Reconstruction Methods on Joint Compensation of Respiratory Motion and Partial Volume Effects in Whole-Body 18F-FDG PET/CT Imaging: Patients with Non-Small Cell Lung Cancer

Background: The present study aims to assess the impact of various image reconstruction methods in 18F-FDG PET/CT imaging on the quantification performance of the proposed technique for joint compensation of respiratory motion and partial volume effects (PVEs) in patients with non-small cell lung cancer. Materials and Methods: An image-based deconvolution technique was proposed, incorporating w...

متن کامل

Optimization of Parameters in 16-slice CT-‌‌scan Protocols for Reduction of the Absorbed Dose

Introduction In computed tomography (CT) technology, an optimal radiation dose can be achieved via changing radiation parameters such as mA, pitch factor, rotation time and tube voltage (kVp) for diagnostic images. Materials and Methods In this study, the brain, abdomen, and thorax scaning was performed using Toshiba 16-slice scannerand standard AAPM and CTDI phantoms. AAPM phantom was used for...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

متن کامل
عنوان ژورنال:
  • Journal of nuclear medicine : official publication, Society of Nuclear Medicine

دوره 53 11  شماره 

صفحات  -

تاریخ انتشار 2012