Beam Characteristics at Low Dose Monitor Unit Settings for Vero4DRT
نویسندگان
چکیده
The stability of delivery of low monitor unit (MU) setting is important especially for step-andshoot intensity-modulated radiotherapy (IMRT), because the nature of the technique is inherent to repeat beam on/off according to the number of the segments. This study evaluates the dose linearity and profile flatness/symmetry under low MU settings for Vero4DRT, a new linear-accelerator based irradiation system that currently implements step-and-shoot IMRT. To evaluate the dose linearity and flatness/symmetry, the point doses and beam profiles were measured as functions of MU and dose rates. The accuracy of dose delivery depended on the dose rate. Under all dose rates, the dose was linear within 1% above 5 MU and within 2% above 3 MU. The beam symmetry was degraded in-line compared with crossline, although both profiles were symmetric within 2% at all dose settings. The profile flatness was also within 2% above 5 MU at any dose rate and showed no significant variation among the low MU settings. To ensure stable beam delivery without increasing the treatment time of Vero4DRT, we recommend a delivery of 5 MU per segment at a dose rate of 500 MU/min.
منابع مشابه
Feasibility evaluation of a new irradiation technique: three-dimensional unicursal irradiation with the Vero4DRT (MHI-TM2000)
The Vero4DRT (MHI-TM2000) is a newly designed unique image-guided radiotherapy system consisting of an O-ring gantry. This system can realize a new irradiation technique in which both the gantry head and O-ring continuously and simultaneously rotate around the inner circumference of the O-ring and the vertical axis of the O-ring, respectively, during irradiation. This technique creates three-di...
متن کاملDevelopment of a dose verification system for Vero4DRT using Monte Carlo method
Vero4DRT is an innovative image-guided radiotherapy system employing a C-band X-ray head with gimbal mechanics. The purposes of this study were to propose specific MC models of the linac head and multileaf collimator (MLC) for the Vero4DRT and to verify their accuracy. For a 6 MV photon beam delivered by the Vero4DRT, a simulation code was implemented using EGSnrc. The linac head model and the ...
متن کاملEvaluation of treatment planning system monitor unit calculations for three intensity modulated radiotherapy delivery techniques
Background: We have validated the monitor unit calculations from a commercially available treatment planning system (TPS) for three intensity modulated radiotherapy (IMRT) planning techniques for tangential breast irradiation by using ionization chamber measurements. Materials and Methods: Treatment plans were generated for forty-two breast patients by a forward planned field in field ...
متن کاملEvaluating Performance of Algorithms in Lung IMRT: A Comparison of Monte Carlo, Pencil Beam, Superposition, Fast Superposition and Convolution Algorithms
Background: Inclusion of inhomogeneity corrections in intensity modulated small fields always makes conformal irradiation of lung tumor very complicated in accurate dose delivery.Objective: In the present study, the performance of five algorithms via Monte Carlo, Pencil Beam, Convolution, Fast Superposition and Superposition were evaluated in lung cancer Intensity Modulated Radiotherapy plannin...
متن کاملDose calibration of EPIDs for segmented IMRT dosimetry
The purpose of this study was to investigate the dose response of amorphous silicon (a-Si) electronic portal imaging devices (EPIDs) under different acquisi- tion settings for both open jaw defined fields and segmented intensity-modulated radiation therapy (IMRT) fields. Four different EPIDs were used. Two Siemens and one Elekta plus a standalone Perkin Elmer research EPID. Each was operated wi...