Magnetic resonance imaging of lung tissue: influence of body positioning, breathing and oxygen inhalation on signal decay using multi-echo gradient-echo sequences.

نویسندگان

  • Andreas Boss
  • Susanne Schaefer
  • Petros Martirosian
  • Claus D Claussen
  • Fritz Schick
  • Juergen F Schaefer
چکیده

PURPOSE To assess susceptibility related signal decay in lung tissue and to measure the influence of body positioning, together with inspiration and expiration, as well as oxygen inhalation. T2* maps and line shape maps of lung parenchyma were derived from datasets acquired at 0.2 T and compared with findings at 1.5 T. The line shape maps allow for a visualization of the intravoxel frequency distribution of lung parenchyma. MATERIALS AND METHODS A multiecho spoiled gradient-echo sequence with 16 echoes was implemented both on a 0.2 T [repetition time (TR) = 100 milliseconds, echo time (TE)1 = 2.15 milliseconds, DeltaTE = 2.94 milliseconds, flip angle 30 degrees] and on a 1.5 T magnetic resonance scanner (TR = 100 milliseconds, TE1 = 1.25 milliseconds, DeltaTE = 1.65 milliseconds, flip angle 30 degrees). Sagittal datasets were recorded in 8 healthy volunteers at 0.2 T in supine position under maximal expiration and inspiration and during oxygen breathing. Additional measurements were performed after 20 minutes inside the scanner in supine position and after prone repositioning. In 2 volunteers, further datasets were acquired at 1.5 T. Color-encoded T2* maps and full-width-at-half-maximum (FWHM) maps of the frequency distribution were computed on a pixel-by-pixel basis. T2* maps were generated by mono-exponential fitting and, additionally, with an extended nonexponential fitting approach. The FWHM maps were calculated with a model-free approach using a discrete Fast Fourier Transformation. RESULTS A notably slower T2* decay was found at 0.2 T (T2*: 5.9-11.8 milliseconds) when compared with 1.5 T (T2*: 1.0-1.4 milliseconds), allowing for the measurement of up to 6 to 8 gradient echoes above the noise level. The T2* maps and the FWHM maps computed from the datasets acquired at 0.2 T allowed regional comparison of the derived parameters. If volunteers were positioned in supine position, expiration resulted in a T2* of 10.9 +/- 1.0 milliseconds and a FWHM of 47.1 +/- 4.0 Hz in the dorsal lung. Significant changes (P < 0.05) were found, eg, in the ventral lung in expiration (T2*: 7.5 +/- 0.8, FWHM: 76.7 +/- 11.2) versus dorsal lung in expiration, in the dorsal lung in inspiration (T2*: 8.4 +/- 1.0, FWHM: 67.8 +/- 12.5) versus dorsal lung in expiration, in the dorsal lung during oxygen breathing (T2*: 8.7 +/- 1.1, FWHM: 52.2 +/- 5.2) versus dorsal lung while breathing room air, and in the dorsal lung in prone position (T2*: 8.5 +/- 0.6, FWHM: 67.0 +/- 9.2) versus dorsal lung in supine position. CONCLUSION The proposed method allows for the computation of color-encoded T2* maps and FWHM maps of lung parenchyma in good image quality using datasets acquired at 0.2 T. The technique is robust and sensitive to physiological changes of lung magnetic resonance properties, eg, due to the type of body positioning or oxygen breathing.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An investigation into the effect of magnetic resonance imaging (MRI) echo time spacing and number of echoes on the sensitivity and dose resolution of PAGATUG polymer-gel dosimeter

Background: There are various methods to read out responses of a polymer-gel dosimeter, among which the Magnetic Resonance Imaging (MRI) technique is the most common one. Optimizing imaging protocols can have significant effect on the sensitivity and the dose resolution of polymer gel dosimeters. This study has investigated the effects of the number of echoes (NOE) and the echo time spacing (ES...

متن کامل

ANALYTICAL STUDY OF EFFECT OF BILAYER INORGANIC AND ORGANIC COATING AROUND THE IRON OXIDE NANOPARTICLES ON MAGNETIC RESONANCE IMAGING CONTRAST

Background & Aims: In recent years, iron oxide nanoparticles have been used in contrast-enhanced magnetic resonance imaging for diagnosing a wide range of diseases. In order to provide biocompatibility and prevent the toxicity of the nanoparticles, using organic or inorganic coating around these nanoparticles is important for their application. The aim of this study is to investigate the effect...

متن کامل

اندازه گیری غیریکنواختی امواج رادیوئی در ام آر آی

Introduction: Non-uniformity is one of the most important parameters affecting MRI images which can lead to harmful effects in the diagnosis and analysis of qualitative and quantitative methods. The present study introduced a method for measuring RF non-homogeneity in MRI systems. Methods and Materials: To verify the uniformity of B0 and B1 fields, a cylindrical phantom with a diameter of 24 c...

متن کامل

Tissue temperature monitoring with multiple gradient-echo imaging sequences.

The inherent sensitivity of multiple gradient-echo sequences to the chemical shift is exploited to rapidly map muscle water frequency shifts caused by ultrasonic heating. The use of multiple echoes is shown to offer several advantages over single gradient-echo approaches previously proposed for temperature measurement. An increase in the effective bandwidth significantly reduces aliasing proble...

متن کامل

Assessment of the Characteristics of MRI Coils in Terms of RF Non-Homogeneity Using Routine Spin Echo Sequences

Introduction: One of the major causes of image non-uniformity in MRI is due to the existence of  non-homogeneity in RF receive and transmit. This can be the most effective source of error in  quantitative  studies  in  MRI  imaging.  Part  of  this  non-homogeneity  demonstrates  the  characteristics of RF coil and part of it is due to the interaction of RF field with the material being  imaged...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Investigative radiology

دوره 43 6  شماره 

صفحات  -

تاریخ انتشار 2008