Albuterol Delivery via Facial and Tracheostomy Route in a Model of a Spontaneously Breathing Child.

نویسندگان

  • Brandy Cooper
  • Ariel Berlinski
چکیده

BACKGROUND Some pediatric patients receiving therapeutic aerosols undergo tracheostomy, and others who are tracheostomized continue requiring inhaled therapies upon decannulation. It is unknown whether a dose adjustment is required. Different devices are available for facial and tracheostomy delivery, and in some instances, the assisted technique is used. We hypothesized that the change from face mask to tracheostomy would result in a decrease in the lung dose. METHODS A breathing simulator connected in series to a filter holder and an anatomically correct head model of a child was used. The drug captured in the filter was termed the lung dose. Breathing patterns with tidal volumes of 50, 155, and 300 mL were tested. Albuterol hydrofluoroalkane (pressurized metered-dose inhaler [pMDI]) with an AeroChamber Mini (face and 4.5-mm tracheostomy), AeroTrach (4.5-mm tracheostomy), and AeroChamber (face) and albuterol (2.5 mg/3 mL) with a continuous output nebulizer (face and 4.5-mm tracheostomy) were tested. Masks were used for facial delivery. Four units of each device were tested. Particle size of the pMDI was measured by cascade impaction. Albuterol concentration was determined via spectrophotometry (276 nm). RESULTS Switching from facial to tracheostomy delivery increased lung dose with nebulizer (all breathing patterns). When a pMDI was used, lung dose was unchanged or increased for the 50- and 155-mL and decreased for the 300-mL breathing pattern. The use of the assisted technique increased lung dose only during nebulization with the 300-mL breathing pattern. The particle size of the pMDI decreased by 19-23% when traveling through the tracheostomy tube, which retained < 4% and < 26% of the nominal dose of a nebulizer and pMDI, respectively. CONCLUSIONS The effect of changing the delivery route from face to tracheostomy was variable and depended on the delivery device and the breathing pattern. There is no advantage of using the assisted technique to enhance aerosol delivery.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Nebulized albuterol delivery in a model of spontaneously breathing children with tracheostomy.

BACKGROUND Nebulized therapy is commonly used in spontaneously breathing tracheostomized patients, despite a lack of recommended devices and techniques. I compared albuterol dose delivered to a model of spontaneously breathing children with tracheostomy, using different nebulizers, tracheostomy tube sizes, inhalation techniques, and breathing patterns. METHODS A tracheostomy model was connect...

متن کامل

Oronasal and Tracheostomy Delivery of Soft Mist and Pressurized Metered-Dose Inhalers With Valved Holding Chamber.

BACKGROUND Some pediatric tracheostomized patients who receive inhaled drugs undergo decannulation, and it is unknown whether the dose has to be adjusted. Pressurized metered-dose inhalers (pMDIs) and soft mist inhalers (SMIs) used with valved holding chambers (VHCs) made of non-electrostatic material are available. We hypothesized that using an SMI and changing the delivery route from tracheos...

متن کامل

The Effect of Aerosol Devices and Administration Techniques on Drug Delivery in a Simulated Spontaneously Breathing Pediatric Model with a Tracheostomy

Background: Evidence on aerosol delivery via tracheostomy is lacking. The purpose of this study was to evaluate the effect of aerosol device and administration technique on drug delivery in a simulated spontaneously breathing pediatric model with tracheostomy. Methods: Delivery efficiencies during spontaneous breathing with assisted and unassisted administration techniques were compared using t...

متن کامل

Quantifying Aerosol Delivery in Simulated Spontaneously Breathing Patients With Tracheostomy Using Different Humidification Systems With or Without Exhaled Humidity.

BACKGROUND Aerosol and humidification therapy are used in long-term airway management of critically ill patients with a tracheostomy. The purpose of this study was to determine delivery efficiency of jet and mesh nebulizers combined with different humidification systems in a model of a spontaneously breathing tracheotomized adult with or without exhaled heated humidity. METHODS An in vitro mo...

متن کامل

Effect of Aerosol Devices and Administration Techniques on Drug Delivery in a Simulated Spontaneously Breathing Pediatric Tracheostomy Model.

BACKGROUND This study was conducted to compare the efficiency of jet nebulizers, vibrating mesh nebulizers, and pressurized metered-dose inhalers (pMDI) during assisted and unassisted administration techniques using a simulated spontaneously breathing pediatric model with a tracheostomy tube (TT). METHODS An in vitro breathing model consisting of an uncuffed TT (4.5-mm inner diameter) was att...

متن کامل