Improved Beam Angle Arrangement in Intensity Modulated Proton Therapy Treatment Planning for Localized Prostate Cancer

نویسندگان

  • Wenhua Cao
  • Gino J. Lim
  • Yupeng Li
  • X. Ronald Zhu
  • Xiaodong Zhang
  • Robert Weiss
چکیده

PURPOSE This study investigates potential gains of an improved beam angle arrangement compared to a conventional fixed gantry setup in intensity modulated proton therapy (IMPT) treatment for localized prostate cancer patients based on a proof of principle study. MATERIALS AND METHODS Three patients with localized prostate cancer retrospectively selected from our institution were studied. For each patient, IMPT plans were designed using two, three and four beam angles, respectively, obtained from a beam angle optimization algorithm. Those plans were then compared with ones using two lateral parallel-opposed beams according to the conventional planning protocol for localized prostate cancer adopted at our institution. RESULTS IMPT plans with two optimized angles achieved significant improvements in rectum sparing and moderate improvements in bladder sparing against those with two lateral angles. Plans with three optimized angles further improved rectum sparing significantly over those two-angle plans, whereas four-angle plans found no advantage over three-angle plans. A possible three-beam class solution for localized prostate patients was suggested and demonstrated with preserved dosimetric benefits because individually optimized three-angle solutions were found sharing a very similar pattern. CONCLUSIONS This study has demonstrated the potential of using an improved beam angle arrangement to better exploit the theoretical dosimetric benefits of proton therapy and provided insights of selecting quality beam angles for localized prostate cancer treatment.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Iterative Approach for Automatic Beam Angle Selection in Intensity Modulated Radiation Therapy Planning

Introduction: Beam-angle optimization (BAO) is a computationally intensive problem for a number of reasons. First, the search space of the solutions is huge, requiring enumeration of all possible beam orientation combinations. For example, when choosing 4 angles out of 36 candidate beam angles, C36 = 58905 possible combinations exist.  Second, any change in a   beam 4 config...

متن کامل

3-Dimensional conformal radiotherapy versus intensity modulated radiotherapy for localized prostate cancer: Dosimetric and radiobiologic analysis

 Background: To analyze the dosimetric and radio biologic advantages between intensity modulated radiotherapy (IMRT) and 3 dimensional conformal radiotherapy (3DCRT) and selection of optimal photon energy for IMRT treatments. Material and methods: 24 patients with localized prostate carcinoma were planned for 3DCRT and IMRT techniques. Radiation dose of 54 Gy with 2 Gy/fraction, was planned to ...

متن کامل

The Relationship between the Different Prostate Intensity Modulated Radiation Therapy Techniques and Patient’s Anatomical Parameters

Introduction: Intensity Modulated Radiotherapy (IMRT) is a common method for treating prostate cancers. Different IMRT methods can apply for irradiating the prostate, but there is not clear that which of these methods can provide better dose distribution for each patient individually. Therefore, the aim of this study was to evaluate the relationship between the different prosta...

متن کامل

Intensity Modulated Proton Therapy as a Boost Treatment after Prostate Seed Implant: A Treatment Planning Study

Purpose: Combination of Prostate Seed Implant (PSI) with External Beam Radiation Therapy (EBRT) remains as an attractive option for patients with intermediate or high-risk prostate cancer. One of the most widely used approaches is to use external beam radiation therapy (EBRT) to deliver boost doses after permanent prostate seed implant (PSI). In this study, the feasibility of using Intensity Mo...

متن کامل

The efficacy of conventional external beam, three-dimensional conformal, intensity-modulated, particle beam radiation, and brachytherapy for localized prostate cancer.

Technologic advances in radiation treatment planning and delivery have generated popular interest in the different radiation therapy techniques used in treating patients with localized prostate cancer. Throughout the past decade, high-energy (> 4 MV) linear accelerators have largely replaced Cobalt machines in external beam radiation therapy (EBRT) delivery. Conventional EBRT has been used to t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 7  شماره 

صفحات  -

تاریخ انتشار 2015