Capillary versus venous bedside blood glucose estimations.
نویسندگان
چکیده
OBJECTIVES To determine the mean difference and correlation between capillary and venous bedside glucose estimation in comparison to laboratory blood glucose analysis in emergency department (ED) patients. METHODS Blood glucose levels were synchronously analysed using a bedside blood glucometer on capillary and venous derived samples from consenting ED patients aged >12 years. The venous sample was sent for comparative testing using a laboratory based multichannel analyser. Mean difference and correlation coefficients were determined. RESULTS A total of 20 subjects (aged 13-88 years) were enrolled, with 100% data capture. The mean laboratory glucose was 7.075 mmol/l. The mean capillary blood glucose was 7.66 mmol/l (mean difference compared with mean laboratory glucose 0.58 mmol/l; 95% confidence interval 0.3 to 0.9). The mean venous derived blood glucometer glucose was 7.99 mmol/l (mean difference compared with mean laboratory glucose 0.91 mmol/l; 95% CI 0.6 to 1.2). The correlation coefficient for the laboratory blood glucose versus the capillary blood glucometer glucose was 0.97 mmol/l (p<0.001). The correlation coefficient for the laboratory blood glucose and the venous blood glucometer glucose was 0.96 (p<0.001). Variation occurred between the glucometer and the laboratory blood glucose results. CONCLUSIONS There is a small but significant difference in the blood glucose results analysed on a bedside glucometer when the samples are taken from capillary or venous sources. Although good correlation is the norm between venous and capillary derived samples, caution must be exercised in accepting the results as equivalent or using either as substitutes for a laboratory blood glucose result.
منابع مشابه
بررسی همخوانی نوارهای سنجش قند خون بتاچک با روش استاندارد
Diabetes mellitus (DM) is the commonest cause of renal failure, blindness and non traumatic amputation. According to ADA, prevention of DM complications is possible through monitoring and control of blood glucose levels.Methods: The Betachek strip, produced in Australia, is able to determine biood glucose level with no need for electronic glucometer devices.Results: In this study, we compared v...
متن کاملAccurate and precise isotope dilution mass spectrometry method for determining glucose in whole blood.
An accurate and precise method to determine glucose concentration in whole blood is presented. The method, based on isotope dilution gas chromatography-mass spectrometry (ID GC-MS), was developed to be used as a Reference Method for determining glucose concentration in capillary or venous whole blood. Blood samples and standards are pipetted manually with "microcap" micropipettes, which makes i...
متن کاملCapillary versus Venous Blood Glucose in Patients with Coma
1 Dear Sir, The recent publication on “comparison of capillary and venous blood glucose levels using glucometer and laboratory blood glucose level in poisoned patients being in coma” is very interesting.[1] Yaraghi et al. concluded that “capillary blood glucose measurement by using glucometer is not recommended for patients in coma.”[1] In fact, there are many previous comparative studies betwe...
متن کاملBedside blood glucose monitoring in hospitals.
The modern management of hospitalized patients with diabetes includes capillary blood glucose determinations at the bedside. This measure is analogous to an additional “vital sign” for people with diabetes. The rapidity with which results can be obtained, and therapeutic decisions made, can improve management and conceivably shorten hospital stays. Replacing venipunctures with capillary blood g...
متن کاملComparability of hemoglobin A1c level measured in capillary versus venous blood sample applying two point-of-care instruments
BACKGROUND The present study is designed to evaluate the validity of the measurement of capillary blood hemoglobin A1c levels in comparison with venous blood hemoglobin A1c. METHODS The data of this cross-sectional study are collected from a sample of 45 Iranian diabetic patients referred to one particular laboratory for the assessment of HbA1c level during a period from April to December 201...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Emergency medicine journal : EMJ
دوره 22 3 شماره
صفحات -
تاریخ انتشار 2005