Delayed diagnosis of catastrophic ceramic liner failure with resultant pelvic discontinuity and massive metallosis

نویسندگان

  • Benjamin A. Winston
  • Ryland P. Kagan
  • Thomas W. Huff
چکیده

With newer-generation ceramic components used in total hip arthroplasty, component fracture is a rare complication. However rare, when ceramic component fracture does occur, prompt identification and revision is necessary as delay can lead to dramatic failure with resultant metallosis as the extremely hard ceramic debris abrades remaining components. We present a case of a 70-year-old woman with ceramic liner fracture and an estimated 10-year delay in intervention with failure resulting in pelvic discontinuity and massive metallosis with associated cutaneous manifestation. She was treated with a complex revision and reconstruction and is 2 years postrevision without major complication.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Catastrophic Failure of the Acetabular Polyethylene Liner in Ceramic-on-Polyethylene Total Hip Arthroplasty

INTRODUCTION Catastrophic polyethylene failure is a rare complication of ceramic-on-polyethylene total hip arthroplasty due to the favorable tribological characteristics of ceramic. Failure of the polyethylene liner can be disastrous, increasing periprosthetic osteolysis, metallosis, and risk of dislocation. Complications associated with ceramic-on-polyethylene articulations have been studied e...

متن کامل

Metallosis after Exchange of the Femoral Head and Liner following Ceramic Acetabular Liner Dissociation in Total Hip Arthroplasty with a Modular Layered Acetabular Component

The type of bearing material that should be used in revision surgery after the failure of ceramic-on-ceramic total hip arthroplasty (THA) remains controversial. In the case of ceramic fracture, the residual ceramic particles can cause consequent metallosis when metal implants are used for revision THA. On the other hand, in the case of THA failure without ceramic fracture, revision THA with a m...

متن کامل

Severe metallosis leading to femoral head perforation.

This article describes a case of severe metallosis in a 67-year-old woman who initially underwent primary total hip arthroplasty with a ceramic-on-ceramic articular bearing. This was subsequently revised to a metal-on-polyethylene articulation due to ceramic liner fracture. She presented with severe hip pain and a pelvic mass. Infective workup was negative. Perforation of the cobalt-chrome femo...

متن کامل

The Result of Revision Total Hip Arthroplasty in Patients with Metallosis Following a Catastrophic Failure of a Polyethylene Liner

BACKGROUND Wear cannot be completely prevented after total hip arthroplasty. If severe polyethylene (PE) liner wear develops, the so-called catastrophic failure occurs and metallosis develops. We postulated that longevity of the new implant may be affected after revision surgery for metallosis following a catastrophic failure of a PE liner due to the substantial amount of PE wear particles and ...

متن کامل

Metallosis: A diagnosis not only in patients with metal-on-metal prostheses

Although the real actual incidence of metallosis is unknown, it is described as a rare diagnosis with a 5% estimated incidence in the hip prosthetic replacements. The adoption of non-metallic articular prosthetic devices, made of polyethylene and ceramic, is the main reason to the diminishing number of reported cases. We present a case of metallosis with a clinical systemic presentation in a pa...

متن کامل