Performance of Single Point Monitors for Measuring Aerial Ammonia and Hydrogen Sulfide
نویسندگان
چکیده
Performance of 43-44 Single Point Monitors (SPMs) was evaluated for measuring aerial ammonia (NH3, 0-30 ppm) and hydrogen sulfide (H2S, 0-90 ppb) under laboratory and field conditions. Calibration gas or sample air with various levels of moisture content or dew-point temperature (tdp) and gas concentrations were introduced simultaneously to the SPMs and a reference gas analyzer – chemiluminescence analyzer for NH3 and pulsed-fluorescence analyzer for H2S. Linear relationships were observed between readings of the SPMs and those of the respective reference analyzer, which were influenced by moisture content or tdp of the calibration or sample air. Specifically, average H2S readings by the SPMs were, respectively, 66%, 80%, 87% and 97% of those by the reference analyzer for calibration gas at tdp of -22 (dry), 9, 13 and 16 oC. In comparison, average NH3 readings by the SPMs were 42%, 86%, 102% and 178% of those by the reference analyzer for calibration gas at tdp of -22, 8.5-10, 12.5-14, and 16-17 oC respectively. Coefficient of variation of “as-is” readings among the tested SPMs was up to 15% for H2S and up to 25% for NH3. Regression equations were developed to compensate for the moisture effect on SPM readings of both gases. The correctional regression equations were able to achieve over 90% of the reference H2S readings; however, such equations were not as effective in predicting or correcting NH3 readings by the SPMs.
منابع مشابه
Performance of Single Point Monitor in Measuring Ammonia and Hydrogen Sulfide Gases
Performance of Single Point Monitors (SPMs) was evaluated for measuring aerial ammonia (NH3, 0 to 30 ppm) and hydrogen sulfide (H2S, 0 to 90 ppb) under laboratory and field conditions. Calibration gas or NH3/H2S−ladden air at various dew−point temperatures (tdp) were introduced simultaneously to the SPMs under evaluation and a chemiluminescence NH3 analyzer or a pulsed−fluorescence H2S analyzer...
متن کاملEffects of aerial ammonia, hydrogen sulfide and swine-house dust on rate of gain and respiratory-tract structure in swine.
Ammonia, hydrogen sulfide and swine-house dust at continuous aerial levels as high as or higher than those normally encountered in enclosed swine-finishing houses had little or no effect on rate of body-weight gain and respiratory-tract structure of 92 healthy pigs in seven trials. The trials were of 17 to 109 days' duration, involved air pollutants alone and in various combinations and were co...
متن کاملComparison of Biofilter Residence Time
Residence time is a critical factor in determining the size of an open face biofilter. Four test biofilters were constructed with media comprised of yard waste compost and brush chips to compare 8 s. and 4 s. residence time. Biofilter performance was monitored for ten months by measuring pressure drop, moisture content, odor threshold, hydrogen sulfide, and ammonia concentrations. Odor reductio...
متن کاملHydrogen sulfide removal from biogas using chemical absorption technique in packed column reactors
The toxicity and corrosion potential of hydrogen sulfide in raw biogas underlines the need for biogas purification. Several techniques available for removal of hydrogen sulfide from biogas are out of the reach for common end users due to lack of knowledge, higher running costs, and insufficient operational skills. The present experimental study aims to propagate hydrogen sulfide removal techniq...
متن کاملMonitoring of aerial pollutants emitted from Swine houses in Korea.
This on-site survey study was performed to determine the concentrations and emissions of aerial contaminants in the different types of swine houses in Korea and then to present beneficial information available for Korean pig producers to manage optimal air quality in swine house. The swine houses investigated in this research were selected based on three criteria; manure removal system, ventila...
متن کامل