Impact of a commercial 3D OSEM reconstruction algorithm on the 177Lu activity quantification of SPECT/ CT imaging in a Molecular Radiotherapy trial

نویسندگان

  • E. Grassi
  • F. Fioroni
  • E. Mezzenga
  • D. Finocchiaro
  • M. A. Sarti
  • A. Filice
  • A. Versari
  • M. Iori
چکیده

The purpose of this study was to investigate the influence of the Ordered Subsets Expectation Maximization (OSEM) reconstruction updates implemented in the 177Lu SPECT/CT imaging processing in molecular radiotherapy. A NEMA IEC Body PhantomTM was used to quantify activity in refillable spheres of five different sizes. Images were obtained with a hybrid dual-head SPECT-CT imaging system (Symbia T2, Siemens Medical System, Germany) with a clinical acquisition protocol, and reconstructed using a commercial 3D OSEM algorithm (Flash 3D). In the reconstruction process, different values of iterations and subsets were considered, along with a 3D Gaussian post-reconstruction filter and scatter and attenuation correction. Activity recovery coefficients were derived from the ratio between total reconstructed counts and the true activity for each sphere at each OSEM update. Recovery coefficients, and average fractional error (i.e. the weighted Root Mean Squared Error) were evaluated. At the same time, also 177Lu spatial resolution and dead time were investigated, as matter of discussion about activity recovery coefficients. Results for spheres ≤ 5.5 ml in volume were significantly affected by the partial volume effect, causing a great bias in activity estimation for the smallest spheres. Their weighted fractional error was OSEM update dependent, ranging between 85% to 79% and 60% to 50% for the two smallest spheres, referring to values of 8 subsets-8 iterations and 16 subsets-10 iterations for the two extremes, respectively. No dead time was detected. The choice of iterations and subsets is dependent on the object size to investigate and on the desired image quality. Anyway, using a fixed number of iterations and subsets is correct for objects with volumes ≥ 5.5 ml, reaching the total count convergence in the reconstructed volumes, but the use of correction factors for compensating the partial volume effect is needed. For objects with volumes ≤ 5.5 ml the quantification becomes challenging. Correspondence to: Dr. Elisa Grassi, Medical Physics Unit, IRCCS-ASMN, Viale Risorgimento 57, 42123 Reggio Emilia, Italy; E-mail: [email protected]

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy

The purpose of this study is to investigate the optimal reference geometry for gamma camera calibration. Yet another question of interest was to assess the influence of the number of 3D Ordered Subsets Expectation Maximization (3D-OSEM) updates on activity quantification for SPECT imaging with 177Lu. The accuracy of 177Lu activity quantification was assessed both in small and in large objects. ...

متن کامل

Collimator-detector response compensation in molecular SPECT reconstruction using STIR framework

Introduction:It is well-recognized that collimator-detector response (CDR) is the main image blurring factor in SPECT.  In this research, we compensated the images for CDR in molecular SPECT by using STIR reconstruction framework. Methods: To assess resolution recovery capability of the STIR, a phantom containing five point sources along with a micro Derenzo p...

متن کامل

Influences of reconstruction and attenuation correction in brain SPECT images obtained by the hybrid SPECT/CT device: evaluation with a 3‐dimensional brain phantom

Objective: The aim of this study was to evaluate the influences of reconstruction and attenuation correction on the differences in the radioactivity distributions in 123I brain SPECT obtained by the hybrid SPECT/CT device. Methods: We used the 3-dimensional (3D) brain phantom, which imitates the precise structure of gray mater, white matter and bone regions. It was filled with 123I solution (20...

متن کامل

Impact of reconstruction method on quantitative parameters of 99mTc-TRODAT-1 SPECT

Introduction: Quantitative evaluation is recommended to improve diagnostic ability and serial assessment of dopamine transporter (DAT) density scans. We decided to compare the ordered subsets expectation-maximization (OSEM) with filtered back-projection (FBP), and to investigate the impact of different iteration and cut-off frequencies on SBR values. Methods</stro...

متن کامل

Impact of Various Image Reconstruction Methods on Joint Compensation of Respiratory Motion and Partial Volume Effects in Whole-Body 18F-FDG PET/CT Imaging: Patients with Non-Small Cell Lung Cancer

Background: The present study aims to assess the impact of various image reconstruction methods in 18F-FDG PET/CT imaging on the quantification performance of the proposed technique for joint compensation of respiratory motion and partial volume effects (PVEs) in patients with non-small cell lung cancer. Materials and Methods: An image-based deconvolution technique was proposed, incorporating w...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2017