Explant analysis of AneuRx stent grafts: relationship between structural findings and clinical outcome.
نویسندگان
چکیده
OBJECTIVE We reviewed the structural findings of explanted AneuRx stent grafts used to treat abdominal aortic aneurysms, and relate the findings to clinical outcome measures. METHODS We reviewed data for all bifurcated AneuRx stent grafts explanted at surgery or autopsy and returned to the manufacturer from the US clinical trial and worldwide experience of more than 33,000 implants from 1996 to 2003. Devices implanted for more than 1 month with structural analysis are included in this article. Explant results were analyzed in relation to cause of explantation and pre-explant evidence of endoleak, enlargement, or device migration. RESULTS One hundred twenty explanted stent grafts, including 37 from the US clinical trial, were analyzed. Mean implant duration was 22 +/- 13 months (range, 1-61 months). Structural abnormalities included stent fatigue fractures, fabric abrasion holes, and suture breaks. The mean number of nitinol stent strut fractures per explanted device was 3 +/- 4, which represents less than 0.2% of the total number of stent struts in each device. The mean number of fabric holes per explanted device was 2 +/- 3, with a median hole size of 0.5 mm(2). Suture breaks were seen in most explanted devices, but composed less than 1.5% of the total number of sutures per device. "For cause" explants (n = 104) had a 10-month longer implant duration (P =.007) compared with "incidental" explants (n = 16). "For cause" explants had more fractures (3 +/- 5; P =.005) and fabric holes (2 +/- 3; P =.008) per device compared with "incidental" explants, but these differences were not significant (P =.3) when adjusted for duration of device implantation. Among clinical trial explants the number of fabric holes in grafts in patients with endoleak (2 +/- 3 per device) was no different from those without endoleak (3 +/- 4 per device; P = NS). The number of fatigue fractures or fabric holes was no different in grafts in clinical trial patients with pre-explant aneurysm enlargement compared with those without enlargement. Pre-explant stent-graft migration was associated with a greater number of stent strut fractures (5 +/- 7 per device; P =.04) and fabric holes (3 +/- 3 per bifurcation; P =.03) compared with explants without migration. Serial imaging studies revealed inadequate proximal, distal, or junctional device fixation as the probable cause of rupture or need for conversion to open surgery in 86% of "for cause" explants. Structural device abnormalities were usually remote from fixation sites, and no causal relationship between device findings and clinical outcome could be established. CONCLUSIONS Nitinol stent fatigue fractures, fabric holes, and suture breaks found in explanted AneuRx stent grafts do not appear to be related to clinical outcome measures. Longer term studies are needed to confirm these observations.
منابع مشابه
Risk-adjusted outcome analysis of endovascular abdominal aortic aneurysm repair in a large population: how do stent-grafts compare?
PURPOSE To compare differences in the applicability and incidence of postoperative adverse events among stent-grafts used for repair of infrarenal aortic aneurysms. METHODS An analysis of 6787 patients from the EUROSTAR Registry database was conducted to compare aneurysm morphological features, patient characteristics, and postoperative events for the AneuRx, EVT/Ancure, Excluder, Stentor, Ta...
متن کاملSuitability of 7 aortic stent-graft models for MRI-based surveillance.
PURPOSE To evaluate the magnetic resonance imaging (MRI) characteristics of commercially available stent-grafts used for abdominal aortic aneurysm repair. METHODS Seven endovascular grafts (AneuRx, Lifepath, Talent, Excluder, Zenith, Quantum LP, and Ancure) were suspended in a water bath containing gadolinium and scanned using a 1.5-T clinical MRI scanner. Two different scan techniques (T(1)-...
متن کاملMathematical Model analysis of Wallstent and Aneurx: dynamic responses of bare-metal endoprosthesis compared with those of stent-graft.
We performed this study in order to analyze the mechanical properties of bare-metal Wallstent endoprostheses and of AneuRx stent-grafts and to compare their responses to hemodynamic forces. Mathematical modeling, numerical simulations, and experimental measurements were used to study the 2 structurally different types of endoprostheses. Our findings revealed that a single bare-metal Wallstent e...
متن کاملThe US AneuRx Clinical Trial: 6-year clinical update 2002.
A total of 1193 patients with infrarenal abdominal aortic aneurysms were treated with the AneuRx Stent Graft System at 19 US investigational centers from 1996 to 1999. This report summarizes clinical data collected and analyzed as of August 28, 2002. There have been 10 late (>30 days) aneurysm ruptures, 8 late (>30 days) aneurysm-related deaths, and 38 late (>30 days) surgical conversions, incl...
متن کاملThe AneuRx stent graft: four-year results and worldwide experience 2000.
OBJECTIVE The objective was to review the current results of endovascular abdominal aortic aneurysm repair with the AneuRx stent graft and to determine the effectiveness of the device in achieving the primary objective of preventing aneurysm rupture. METHODS The outcome of all patients treated during the past 4 years in the U.S. AneuRx clinical trial was determined, and the worldwide clinical...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of vascular surgery
دوره 40 1 شماره
صفحات -
تاریخ انتشار 2004