Accuracy of noninvasive and continuous hemoglobin measurement by pulse co-oximetry during preoperative phlebotomy.

نویسندگان

  • Elisabeth Dewhirst
  • Aymen Naguib
  • Peter Winch
  • Julie Rice
  • Mark Galantowicz
  • Patrick McConnell
  • Joseph D Tobias
چکیده

BACKGROUND In recent years, the continuous noninvasive hemoglobin measurement has been offered by devices using advanced pulse oximetry technology. Accuracy has been established in healthy adults as well as in surgical and intensive care unit patients but not in the setting of acute hemorrhage. In this study, we evaluated the accuracy of such a device in the clinical setting of preoperative phlebotomy thereby mimicking a scenario of acute blood loss. METHODS This prospective study included patients undergoing surgical repair of congenital heart disease (CHD) for whom preoperative phlebotomy was planned. Blood was removed after the induction of anesthesia and prior to the start of the surgical procedure. Replacement with crystalloid was guided by hemodynamic variables and cerebral oxygenation measured by near-infrared spectroscopy. Hemoglobin was measured by bedside whole blood analysis (total hemoglobin [tHb]) before and after phlebotomy, and concurrent measurements from the pulse co-oximeter (noninvasive, continuous, or spot-check testing of total hemoglobin [SpHb]) were recorded. RESULTS The study cohort included 45 patients ranging in age from 3 months to 50 years. Preoperative phlebotomy removed an average of 9.2 mL/kg of blood that was replaced with an average of 7.2 mL/kg of crystalloid. The pre- and postphlebotomy tHb values were 13.0 ± 1.9 and 12.4 ± 1.8 g/dL, respectively. The absolute difference between the tHb and SpHb (▵Hb) was 1.2 ± 0.1 g/dL. Bland-Altman analysis revealed a bias of 0.1 g/dL, a precision of 1.5 g/dL, and 95% limits of agreement of -2.8 to 3.1 g/dL. In 52.2% of the sample sets, the SpHb was within 1 g/dL of the actual hemoglobin value (tHb), and in 80% of the sample sets, the SpHb was within 2 g/dL. No variation in the accuracy of the deviation was noted based on the patient's age, weight, or type of CHD (cyanotic versus acyanotic). CONCLUSION The current study demonstrates that the accuracy of continuous, noninvasive hemoglobin measurement was not affected by acute blood loss simulated by preoperative phlebotomy. Although the device provided a clinically acceptable correlation with the actual hemoglobin value and offers the value of a continuous trend monitor, given the precision of the device, it does not appear that actual transfusion decisions can be based on the device alone.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients.

OBJECTIVE To determine whether noninvasive hemoglobin measurement by Pulse CO-Oximetry could provide clinically acceptable absolute and trend accuracy in critically ill patients, compared to other invasive methods of hemoglobin assessment available at bedside and the gold standard, the laboratory analyzer. DESIGN Prospective study. SETTING Surgical intensive care unit of a university teachi...

متن کامل

Trending, Accuracy, and Precision of Noninvasive Hemoglobin Monitoring During Human Hemorrhage and Fixed Crystalloid Bolus.

Automated critical care systems for en route care will rely heavily on noninvasive continuous monitoring. It has been reported that noninvasive assessment of blood hemoglobin via CO-oximetry (SpHb) assessed by spot measurements lacks sufficient accuracy for clinical decision making in trauma patients. However, the precision and utility of trending of continuous hemoglobin have not been evaluate...

متن کامل

The efficacy of noninvasive hemoglobin measurement by pulse CO-oximetry in neonates.

OBJECTIVE To evaluate clinical applicability of noninvasive hemoglobin (Hb) measurement with a pulse CO-oximeter in neonates. DESIGN Prospective comparison study. SETTING Neonatal ICU. PATIENTS Fifty-six preterm and term infants with median age = 20 days (range = 1-98 days) and median weight = 1,440 g (range = 530-4,230 g). INTERVENTIONS Hb measurements by Pulse CO-Oximetry (Masimo Radi...

متن کامل

Accuracy of noninvasive haemoglobin measurement by pulse oximetry depends on the type of infusion fluid

Context Measurement of blood haemoglobin (Hb) concentration by pulse oximetry (SpHb) could be of value to determine when erythrocytes should be transfused during surgery, but the effect of infusion fluids on the results is unclear. Objective To study the effect of crystalloid and colloid fluid on the accuracy (bias) and precision of SpHb to indicate the venous Hb concentration in volunteers. De...

متن کامل

Effects of cardiopulmonary bypass on the accuracy of non-invasive hemoglobin measurement by pulse co-oximetry

Introduction A non-invasive continuous measurement of hemoglobin (Hb) by pulse co-oximetry (SpHb) using Radical-7 (Masimo Corporation, CA, USA) is a newly developed method of measuring total Hb concentration. Previously, the accuracy of SpHb in clinical settings was reported (1). However, it has been also reported that SpHb has poor correlation with arterial Hb after cardiac surgery (2). Here w...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of intensive care medicine

دوره 29 4  شماره 

صفحات  -

تاریخ انتشار 2014