Shear force at failure and stiffness of all-inside meniscal repair devices.

نویسندگان

  • William K Kesto
  • Amanda O Esquivel
  • David C Markel
چکیده

The purpose of this study was to determine the failure load and stiffness of various meniscal repair devices. A total of 61 fresh-frozen porcine menisci (medial and lateral) were used for the study. A 30-mm vertical, full-thickness tear was created and repaired using one of three all-inside fixation devices and one inside-out repair in the vertical mattress pattern. We used the MaxBraid (Biomet, Warsaw, IN) inside-out suture as a control. The other devices tested were the Meniscal Cinch (Arthrex, Naples, FL), Ultra FasT-Fix (Smith & Nephew, Andover, MA), and the MaxFire MarXmen (Biomet, Warsaw, IN). In addition, two devices, MaxFire MarXmen and Ultra FasT-Fix, were tested using a horizontal mattress configuration. Using the vertical mattress pattern, the Meniscal Cinch had the highest average load to failure. The Meniscal Cinch was significantly less stiff than the other three devices (p < 0.04). For the MarXmen and Ultra FasT-Fix, no differences were noted for load to failure between horizontal and vertical mattress patterns. The mode of failure was significantly different when comparing the two different surgical techniques for the MaxFire MarXmen (p = 0.005). The MaxFire MarXmen device produced a significantly stiffer (p < 0.001) construct when following the manufacturer's instructions (5.8 N/mm) than with the technique used for the other all-inside devices (2.5 N/mm) The Meniscal Cinch had the highest load-to-failure value but the lowest stiffness of the group in the vertical mattress configuration. There was little difference in biomechanical properties between vertical and horizontal repair. Importantly, there was a significant difference in stiffness and failure mode for the MaxFire MarXmen when the manufacturer guidelines were not specifically followed.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Tensile and shear loading stability of all-inside meniscal repairs: an in vitro biomechanical evaluation.

BACKGROUND Most biomechanical studies for evaluation of the structural properties of meniscal repairs have been performed in tensile loading scenarios perpendicular to the circumferential meniscal fibers. However, meniscal repair constructs are also exposed to shear forces parallel to the circumferential meniscal fibers during healing, particularly in the midportion of the meniscus. HYPOTHESI...

متن کامل

All-inside meniscus repair: a systematic review.

All-inside meniscus repairs are becoming more common with the invention of meniscal arrows, darts, screws, staples, and other suture devices. With the rising popularity of these new all-inside meniscal repair devices, it is necessary to assess the effectiveness of various all-inside surgical interventions used in the treatment of meniscal tears. We performed an evidence based review of the outc...

متن کامل

Elongation and Structural Properties of Meniscus Repair Using Suture Techniques in Distraction and Shear Force Scenarios

Introduction: Meniscal debridement remains the most commonly performed orthopaedic surgery in the US, even though several clinical long term studies have shown that total or partial meniscectomy may lead to cartilage degeneration and early onset of osteoarthritic changes. Repair should be considered depending on the type and the location of the meniscal tear. Good and excellent results have bee...

متن کامل

A biomechanical evaluation of all-inside 2-stitch meniscal repair devices with matched inside-out suture repair.

BACKGROUND Many all-inside suture-based devices are currently available, including the Meniscal Cinch, FasT-Fix, Ultra FasT-Fix, RapidLoc, MaxFire, and CrossFix System. These different devices have been compared in various configurations, but to our knowledge, the Sequent meniscal repair device, which applies running sutures, has not been compared with the Ultra FasT-Fix, nor has it been compar...

متن کامل

Arthroscopic Meniscal Repair: “Modified Outside-In Technique”

Background:  Despite the introduction of different techniques for meniscal repair, no single procedure is superior in all situations. The new method for meniscal repair named “ modified outside-in technique ” aims to achieve higher primary fixation strength by an alternative suture technique as well as avoid disadvantages of outside-in, inside-out, and all-inside suture procedures. Additionally...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The journal of knee surgery

دوره 26 6  شماره 

صفحات  -

تاریخ انتشار 2013