Validity of Arm-to-Arm BIA Devices Compared to DXA for Estimating % Fat in College Men and Women

نویسندگان

  • REBECCA A. ROCKAMANN
  • EMILY K. DALTON
  • JANA L. ARABAS
  • LIZ JORN
  • JERRY L. MAYHEW
چکیده

Bioelectric impedance analysis (BIA) devices are commonly used to estimate percent body fat (%fat), although validation of their accuracy varies widely. The purpose of this study was to assess the validity of four commonly used BIA devices compared to dual-energy X-ray absorptiometry (DXA). College-aged men (n = 29, age = 19.7 ± 1.2 y, weight = 76.9 ± 12.5 kg) and women (n = 31, age = 20.5 ± 0.8 y, weight = 61.5 ± 9.2 kg) were evaluated for %fat using four single-frequency (50 mHz) BIA devices and DXA. A gender × device repeated measures ANOVA indicated some less expensive BIA devices produced %fat values that were not significantly different from DXA. A thumb-to-thumb BIA device produced the closest values in men (21.9 ± 6.6%) and women (32.1 ± 5.3%) compared to DXA (20.6 ± 6.1% and 30.3 ± 5.4%, respectively). The two more expensive BIA devices significantly underestimated in men (14.7 ± 5.8% and 17.0 ± 5.6%) and women (23.3 ± 4.2% and 23.3 ± 4.2%) compared to DXA. Interclass correlation coefficients with DXA were higher for the more expensive devices in men (ICC = 0.899 and 0.958) than the less expensive devices (ICC = 0.681 and 0.730). In women, all BIA devices showed moderate correlations with DXA (ICC = 0.537 to 0.658). Despite the convenience of simple BIA devices, their use in estimating body composition in young men and women might be questionable due to large variations in the differences between DXA and each device in this study.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessing body composition with DXA and bioimpedance: effects of obesity, physical activity, and age.

OBJECTIVE This study evaluated to what extent dual-energy X-ray absorptiometry (DXA) and two types of bioimpedance analysis (BIA) yield similar results for body fat mass (FM) in men and women with different levels of obesity and physical activity (PA). METHODS AND PROCEDURES The study population consisted of 37-81-year-old Finnish people (82 men and 86 women). FM% was estimated using DXA (GE ...

متن کامل

Validity of segmental multiple-frequency bioelectrical impedance analysis to estimate body composition of adults across a range of body mass indexes.

OBJECTIVE We compared body composition estimates using an eight-electrode, segmental, multiple-frequency bioelectrical impedance analysis (segmental MF-BIA) and dual x-ray absorptiometry (DXA) in a group of healthy adults with a range of body mass indexes (BMIs). METHODS Percentage of body fat (%BF), fat-free mass, and fat mass assessed by DXA and segmental MF-BIA in 132 healthy adults were c...

متن کامل

Comparison of total body potassium with other techniques for measuring lean body mass in men and women with AIDS wasting.

BACKGROUND Lean body mass is an important predictor of survival and functional status in patients with AIDS wasting. The bias between different techniques for assessing body composition in AIDS wasting is not known. DESIGN We compared total body potassium (TBK) with fat-free mass (FFM) determined by dual-energy X-ray absorptiometry (DXA), bioelectrical impedance analysis (BIA), and skinfold-t...

متن کامل

Modeling upper and lower limb muscle volume by bioelectrical impedance analysis.

Most studies employing bioelectrical impedance analysis (BIA) for estimating appendicular skeletal muscle mass using descriptive BIA models rely on statistical rather than biophysical principles. The aim of the present study was to evaluate the feasibility of estimating arm and leg muscle volume (MV) based on multiple bioimpedance measurements and using a recently proposed mathematical model an...

متن کامل

Comparison of multifrequency bioelectrical impedance analysis with dual-energy X-ray absorptiometry for assessment of percentage body fat in a large, healthy population.

BACKGROUND Bioelectrical impedance analysis (BIA) is widely used in clinics and research to measure body composition. However, the results of BIA validation with reference methods are contradictory, and few data are available on the influence of adiposity on the measurement of body composition by BIA. OBJECTIVE The goal was to determine the effects of sex and adiposity on the difference in pe...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 10  شماره 

صفحات  -

تاریخ انتشار 2017