Assessment of central corneal thickness and corneal endothelial morphology using ultrasound pachymetry, non-contact specular microscopy, and Confoscan 4 confocal microscopy
نویسندگان
چکیده
BACKGROUND The aim was to compare the repeatability, reproducibility and inherent precision of ultrasound pachymetry (USP), noncontact specular microscopy (SP-2000P) and the Confoscan 4 confocal microscope (z-ring CS4) in measuring endothelial cell density (ECD), coefficient of variation of cell size (CV), and central corneal thickness (CCT) in normal eyes. METHODS In this prospective study, one eye was selected from each of 30 subjects for the measurements of ECD, CV and CCT, which were taken by two observers. Results were analyzed statistically by repeated-measures analysis of variance (ANOVA) for intra-observer repeatability, inter-observer reproducibility, unpaired t-test, paired t-test, and Bland-Altman analyses to determine limits of agreement (LOA) between the three instruments. RESULTS Mean ECD, measured by SP-2000P and z-ring CS4, were 3115.50 ± 279.70 cells/mm(2) and 3167.50 ± 264.75 cells/mm(2), respectively (observer 1), and 3192.63 ± 249.42 cells/mm(2) (z-ring, observer 2). Mean CV measurements were 27.12 ± 2.51 and 27.10 ± 2.41 (SP-2000P and z-ring CS4, respectively; observer 1), and 27.17 ± 2.25 (z-ring, observer 2). Mean CCT values were 555.11 ± 35.83 μm (USP), 535.82 ± 41.10 μm (SP-2000P) and 552.57 ± 36.83 μm (z-ring CS4), and 554.97 ± 36.34 μm (z-ring CS4, observer 2). However, pairwise tests in all cases there was good repeatability and reproducibility as shown by inter-observer and intra-observer analysis of variance for each of the instruments. CONCLUSIONS The SP-2000P and the z-ring CS4 can be used interchangeably to measure ECD and CV. For CCT, the sample size was too small to test for differences of the CCT measurements between the three instruments.
منابع مشابه
Cigarette Smoking and Pachymetry and Specular Microscopy Parameters of Cornea; A Systematic Review and Meta-analysis
Purpose: To investigate the effect of smoking on corneal parameters, including Endothelial Cell Density, Central Corneal Thickness, Percent of Hexogonality, and Coefficient of Variation. Methods: This study was a systematic review and meta-analysis of published articles until the end of February 2019. To do this, the first search was done for creditable medical databases such as PubMed, and ISI...
متن کاملComparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis.
AIMS To compare central corneal thickness measurements of three pachymetry devices in eyes after laser in situ keratomileusis (LASIK). METHODS Central corneal thickness was measured in 203 eyes after myopic LASIK. Orbscan II scanning slit topography (Bausch & Lomb), SP-2000P non-contact specular microscopy (Topcon), and ultrasonic pachymetry (Tomey) were used in this sequence. RESULTS Three...
متن کاملSCIENTIFIC REPORT Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis
Aims: To compare central corneal thickness measurements of three pachymetry devices in eyes after laser in situ keratomileusis (LASIK). Methods: Central corneal thickness was measured in 203 eyes after myopic LASIK. Orbscan II scanning slit topography (Bausch & Lomb), SP-2000P non-contact specular microscopy (Topcon), and ultrasonic pachymetry (Tomey) were used in this sequence. Results: Three ...
متن کاملCentral Corneal Thickness and Corneal Endothelial Cell Changes Caused by Contact Lens Use in Diabetic Patients
PURPOSE To analyze the effects of soft contact lenses on central corneal thickness and morphologic characteristics of the corneal endothelium in diabetic patients. MATERIALS AND METHODS Ultrasound pachymetry and noncontact specular microscopy were performed on 26 diabetic patients who regularly use soft contact lenses (group 1), 27 diabetic patients who do not use soft contact lenses (group 2...
متن کاملCentral corneal thickness: z-ring corneal confocal microscopy versus ultrasound pachymetry.
PURPOSE To compare the repeatability and validity of corneal pachymetry by a corneal confocal microscope with a z-axis adapter (Confoscan 4.0 with z-ring adapter: z-CS4) versus ultrasound (US) pachymetry in the measurement of central corneal thickness (CCT). METHODS CCT in 44 eyes of 44 subjects was determined with z-CS4. Z-CS4 exams were used to estimate the repeatability of thickness measur...