Body Fat Mass Assessment: A Comparison between an Ultrasound-Based Device and a Discovery A Model of DXA

نویسندگان

  • Jean-Claude Pineau
  • Loïc Lalys
  • Massimo Pellegrini
  • Nino Carlo Battistini
چکیده

Objective. To examine measurement of body composition by ultrasound compared with a reference technique:dual energy X-ray absorptiometry (DXA). We evaluated the accuracy of a portable ultrasound-based device in estimating total body fat mass with those assessed by DXA in adult. Methods. Body fat mass has been estimated using a portable ultrasound-based device in comparison with a contemporary reference DXA apparatus: the Hologic Discovery A. Anthropometric data has been assessed in order to maximize the output of the software associated with the ultrasound-based device. A cross-validation between ultrasound technique (US) and DXA was developed in this study. Total body fat mass estimated by ultrasound was compared with this DXA model in a sample of 83 women and 41 men. Results. Ultrasound technique (US) of body fat (BF) was better correlated with DXA in both women (r (2) = 0.97, P < 0.01) and men (r (2) = 0.92, P < 0.01) with standard errors of estimates (SEE) being 2.1 kg and 2.2 kg, respectively. Conclusion. The use of a portable device based on a US produced a very accurate BF estimate in relation to DXA reference technique. As DXA absorptiometry techniques are not interchangeable, the use of our ultrasound-based device needs to be recalibrated on a more contemporary DXA.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Relationships among dual-energy X-ray absorptiometry, bioelectrical impedance and ultrasound measurements of body composition of swine

In three separate studies (156 pigs total), dual-energy X-ray absorptiometry (DXA), bioelectrical impedance (BIA) and ultrasound were compared as methods for measuring live body composition of pigs at 60 and 100-110 kg BWt. DXA measured total body fat and lean content, BIA measurements of resistance (Rs) and reactance (Xc) were used to calculate total body lean mass and ultrasound measurements ...

متن کامل

Comparison of Fat Distribution Parameters between Individuals with and without Metabolic Syndrome

Abstract Objective: The aim of this study was to compare the body fat parameters in individuals with and without metabolic syndrome using an objective method. Materials and methods: This case-control study was performed in Tehran and consisted of 105 individuals with age range of 20 to 55 years. All participants were overweight and obese. NCEP ATP III criteria were used for identifying subjec...

متن کامل

DXA: can it be used as a criterion reference for body fat measurements in children?

OBJECTIVE Dual-energy X-ray absorptiometry (DXA) is often cited as a criterion method for body composition measurements. We have previously shown that a new DXA software version (Hologic Discovery V12.1) will affect whole-body bone mineral results for subjects weighing <40 kg. We wished to reanalyze pediatric whole-body scans in order to assess the impact of the new software on pediatric soft-t...

متن کامل

AENDO September 40/3

Proctor, D. N., P. C. O’Brien, E. J. Atkinson, and K. S. Nair. Comparison of techniques to estimate total body skeletal muscle mass in people of different age groups. Am. J. Physiol. 277 (Endocrinol. Metab. 40): E489–E495, 1999.—An estimate of total body muscle mass with dual-energy X-ray absorptiometry (DXA; appendicular muscle mass divided by 0.75) was compared with 24-h urinary creatinine ex...

متن کامل

Comparison of techniques to estimate total body skeletal muscle mass in people of different age groups.

An estimate of total body muscle mass with dual-energy X-ray absorptiometry (DXA; appendicular muscle mass divided by 0.75) was compared with 24-h urinary creatinine excretion in 59 healthy men and women [20-30 yr (younger), 45-59 yr (middle age), and 60-79 yr (older)] who stayed in a clinical research center for 5 days. Total body water (2H2O dilution), fat (underwater weighing), bone mineral ...

متن کامل