Validation of an automated solid-phase extraction method for the analysis of 23 opioids, cocaine, and metabolites in urine with ultra-performance liquid chromatography-tandem mass spectrometry.
نویسندگان
چکیده
The aim of this work was to automate a sample preparation procedure extracting morphine, hydromorphone, oxymorphone, norcodeine, codeine, dihydrocodeine, oxycodone, 6-monoacetyl-morphine, hydrocodone, ethylmorphine, benzoylecgonine, cocaine, cocaethylene, tramadol, meperidine, pentazocine, fentanyl, norfentanyl, buprenorphine, norbuprenorphine, propoxyphene, methadone and 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine from urine samples. Samples were extracted by solid-phase extraction (SPE) with cation exchange cartridges using a TECAN Freedom Evo 100 base robotic system, including a hydrolysis step previous extraction when required. Block modules were carefully selected in order to use the same consumable material as in manual procedures to reduce cost and/or manual sample transfers. Moreover, the present configuration included pressure monitoring pipetting increasing pipetting accuracy and detecting sampling errors. The compounds were then separated in a chromatographic run of 9 min using a BEH Phenyl analytical column on a ultra-performance liquid chromatography-tandem mass spectrometry system. Optimization of the SPE was performed with different wash conditions and elution solvents. Intra- and inter-day relative standard deviations (RSDs) were within ±15% and bias was within ±15% for most of the compounds. Recovery was >69% (RSD < 11%) and matrix effects ranged from 1 to 26% when compensated with the internal standard. The limits of quantification ranged from 3 to 25 ng/mL depending on the compound. No cross-contamination in the automated SPE system was observed. The extracted samples were stable for 72 h in the autosampler (4°C). This method was applied to authentic samples (from forensic and toxicology cases) and to proficiency testing schemes containing cocaine, heroin, buprenorphine and methadone, offering fast and reliable results. Automation resulted in improved precision and accuracy, and a minimum operator intervention, leading to safer sample handling and less time-consuming procedures.
منابع مشابه
Relative Determination Approach to the Metabolites of Protoberberine Alkaloids in Rat Urine by Liquid Chromatography Tandem Mass Spectrometry for the Comparative Studies on Rhizome coptidis and Zuojinwan Preparation
The lack of authentic standards has limited the quantitative analysis of herbal drugs in biological samples. The present work demonstrated a practicable strategy for the assay of herbs and their metabolites independent of authentic standards. A liquid chromatography–electrospray ionization–mass spectrometry (LC–ESI–MS) method for the qualitative and quantitative determination of the metabolites...
متن کاملRelative Determination Approach to the Metabolites of Protoberberine Alkaloids in Rat Urine by Liquid Chromatography Tandem Mass Spectrometry for the Comparative Studies on Rhizome coptidis and Zuojinwan Preparation
The lack of authentic standards has limited the quantitative analysis of herbal drugs in biological samples. The present work demonstrated a practicable strategy for the assay of herbs and their metabolites independent of authentic standards. A liquid chromatography–electrospray ionization–mass spectrometry (LC–ESI–MS) method for the qualitative and quantitative determination of the metabolites...
متن کاملDevelopment and Validation of Bioanalytical Method for Simultaneous Estimation of Nebivolol Enantiomers in Human Plasma Using Liquid Chromatography-tandem Mass Spectrometry
The present study describes a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the simultaneous determination of S-RRR and R-SSS nebivolol (nebivolol enantiomers) in human plasma using solid phase extraction technique. Method of both S-RRR and R-SSS nebivolol (nebivolol enantiomers) has been developed and validated using racemic nebivolol D4 as an internal standard. Analytes...
متن کاملComprehensive Automation of the SPE-GC/MS Analysis of Opioids, Cocaine and Metabolites from Serum and Other Matrices
Analyzing blood serum for opioids, cocaine and metabolites is a routine task in forensic laboratories. The most commonly used methods involve several manual or partly-automated sample preparation steps such as protein precipitation, solid phase extraction, evaporation and derivatization followed by GC/MS or LC/MS determination. In this study a comprehensively automated method is compared with a...
متن کاملComprehensive automation of the solid phase extraction gas chromatographic mass spectrometric analysis (SPE-GC/MS) of opioids, cocaine, and metabolites from serum and other matrices
The analysis of opioids, cocaine, and metabolites from blood serum is a routine task in forensic laboratories. Commonly, the employed methods include many manual or partly automated steps like protein precipitation, dilution, solid phase extraction, evaporation, and derivatization preceding a gas chromatography (GC)/mass spectrometry (MS) or liquid chromatography (LC)/MS analysis. In this study...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of analytical toxicology
دوره 38 5 شماره
صفحات -
تاریخ انتشار 2014