A comprehensive EPID-based 3D validation technique for TrueBeam-delivered VMAT plans
نویسندگان
چکیده
Purpose: To develop and validate a pre-treatment EPI dosimetry method on Varian TrueBeam linacs using continuous imaging, with reconstruction in a 3D cylindrical phantom geometry. Methods: Delivery of VMAT plans with continuous imaging is currently possible only in Research Mode on TrueBeam linacs, with images acquired in a proprietary format. An earlier technique was adapted to take advantage of technical improvements in EPID delivery, and was tested under various acquisition conditions. The dosimetry of VMAT plans was evaluated at isocentre and within patient volumes that had been transferred to the virtual phantom. Results: Approximately 60 portal image projections per arc were found to be adequate for 3D reconstruction in phantom volumes of 28cm diameter. Twelve prostate, CNS and Head & Neck deliveries were evaluated in Research mode relative to the corresponding Eclipse (v.10) treatment plans, and to measurements on an ArcCheck device in Treatment mode. Mean dose differences at isocentre were within 2% for the three-way comparison, and in PTV volumes were within 1% (s.d. 1%). However, some discrepancies were observed in ArcCheck results that may be related to the small dimensions of certain VMAT apertures. Conclusions: EPI dosimetry with 3D dose reconstruction is an accurate, comprehensive and efficient pre-treatment validation technique for VMAT delivery. Although currently limited to a research mode on TrueBeam, it has the potential to be implemented for clinical use.
منابع مشابه
Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT
Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...
متن کاملA Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images
The aim in this study is to develop a generalized strategy for 3D dose verification of IMRT and VMAT planes using EPID transit images in combination with Monte Carlo (MC) simulations. An EPID-based dosimetric verification procedure was developed to convert EPID-measured transit images into 2D exit photon fluence by de-convoluting with the MC-simulated EPID response kernels. The present scatter ...
متن کاملAssessment of a 2D EPID-based Dosimetry Algorithm for Pre-treatment and In-vivo Midplane Dose Verification
Introduction: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans both pretreatment and in-vivo. The aim of this study was to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropomorphic phantom and in-vivo, as well. Materials and Methods: </strong...
متن کاملSU-E-T-164: Clinical Implementation of ASi EPID Panels for QA of IMRT/VMAT Plans.
PURPOSE To investigate various issues for clinical implementation of aSi EPID panels for IMRT/VMAT QA. METHODS Six linacs are used in our clinic for EPID-based plan QA; two Varian Truebeams, two Varian 2100 series, two Elekta Infiniti series. Multiple corrections must be accounted for in the calibration of each panel for dosimetric use. Varian aSi panels are calibrated with standard dark fiel...
متن کاملPractical approach for pretreatment verification of IMRT with flattening filter‐free (FFF) beams using Varian portal dosimetry
Patient-specific pretreatment verification of intensity-modulated radiation therapy (IMRT) or volumetric-modulated arc therapy (VMAT) is strongly recommended for all patients in order to detect any potential errors in treatment planning process and machine deliverability, and is thus performed routinely in many clinics. Portal dosimetry is an effective method for this purpose because of its pro...
متن کامل