Lung EIT: Should we reconstruct resistivity or conductivity?
نویسندگان
چکیده
MOTIVATION: In lung EIT, the quantity of interest is the change in regional air volume in the lungs. However, the actual quantity reconstructed is change in a bulk electrical property, usually conductivity, which is not always easy to interpret. In this study, we investigate how the choice of the reconstructed physical quantity affects the properties of the sensitivity matrix and images obtained with linear reconstruction algorithms. METHODS: We simulate EIT measurements on a 3D finite element model with 16 electrodes in a plane. The model contains two non-conductive objects that represent the lungs. Assuming that lung resistivity is linearly proportional to air volume [1], we divide the lungs into an upper and lower part whose aeration we manipulate. We then reconstruct a difference measurement representing a fixed total tidal volume (0.5 L) distributed between the upper and lower lung area as conductivity (σ), log conductivity (log σ) or resistivity (ρ). We present results of 3 simulations with different initial lung aeration and distribution of tidal volume (Figure 1, left), roughly corresponding to a recruitment maneuver. We use the NOSER [2] approach to linear image reconstruction, as implemented in EIDORS, where the sensitivity matrix is calculated either on a homogeneous model or one reflecting the actual initial aeration. RESULTS: Sensitivity (measurement response to unit change in a single element) varies greatly depending on the physical quantity (unit) reconstructed (Figure 1, right). Sensitivity to conductivity change is highest in non-conductive lung, while the opposite holds for resistivity and (to lesser degree) log conductivity. Reconstructions using homogeneous background (Figure 2, left) are not affected by the choice of physical quantity and show increased sensitivity to changes in conductive (less aerated) lung areas. In case 3, the uniformly distributed tidal volume appears concentrated in the lower part of the image. When a background is used (Figure 2, right), pronounced differences between the three approaches emerge. Resistivity reconstructions recover the tidal volume distribution accurately, while log conductivity and conductivity reconstructions remain skewed towards conductive lung areas. They also show a variability in image amplitude not reflecting changes in total tidal volume. CONCLUSION: Lung EIT images reconstructed without regard for the non-homogeneous distribution of conductivity in the thorax are a poor reflection of the distribution of tidal volume changes. When a realistic background conductivity is used, more accurate images can be obtained, depending on the choice of reconstructed quantity and reconstruction algorithm.
منابع مشابه
Choice of reconstructed tissue properties affects interpretation of lung EIT images.
Electrical impedance tomography (EIT) estimates an image of change in electrical properties within a body from stimulations and measurements at surface electrodes. There is significant interest in EIT as a tool to monitor and guide ventilation therapy in mechanically ventilated patients. In lung EIT, the EIT inverse problem is commonly linearized and only changes in electrical properties are re...
متن کاملImplementation of Radon Transformation for Electrical Impedance Tomography (EIT)
Radon Transformation is generally used to construct optical image (like CT image) from the projection data in biomedical imaging. In this paper, the concept of Radon Transformation is implemented to reconstruct Electrical Impedance Topographic Image (conductivity or resistivity distribution) of a circular subject. A parallel resistance model of a subject is proposed for Electrical Impedance Top...
متن کاملLevel-set-based reconstruction algorithm for EIT lung images: first clinical results.
We show the first clinical results using the level-set-based reconstruction algorithm for electrical impedance tomography (EIT) data. The level-set-based reconstruction method (LSRM) allows the reconstruction of non-smooth interfaces between image regions, which are typically smoothed by traditional voxel-based reconstruction methods (VBRMs). We develop a time difference formulation of the LSRM...
متن کاملComment on Borges et al. "regional lung perfusion estimated by electrical impedance tomography in a piglet model of lung collapse".
TO THE EDITOR: The paper by Borges et al. (1) presents an excellent experimental study comparing two approaches for the evaluation of regional pulmonary perfusion by electrical impedance tomography (EIT) with single photon emission computerized tomography (SPECT) as well established reference method. The negative results on the pulsatility approach confirm the skepticism addressed by recently p...
متن کاملNeural-fuzzy Modelling of Lung Volume using Absolute Electrical Impedance Tomography
Electrical Impedance Tomography (EIT) has been the subject of intensive research since its development in the early 1980s by Barber and Brown at the Department of Medical Physics and Clinical Engineering, Hallamshire Hospital in Sheffield (UK). In particular, pulmonary measurement has been the focus of most EIT related research. One of the relatively recent advances in EIT is the development of...
متن کامل