Cross contamination of blood cultures associated with a multiple use venting device.
نویسندگان
چکیده
AIM To investigate cross contamination of blood cultures associated with multiple use venting devices, which are widely used in clinical microbiology laboratories to reduce labour costs. METHODS Systematic analysis of 13,880 blood culture results in a large teaching hospital where multiple use venting devices were employed. RESULTS Nine series of potential cross contamination were identified in a 12 month period. Four series involved coagulase negative staphylococci and were unlikely to represent true cross contamination. Five series involved blood cultures which had significant bacterial growth at the time of venting. CONCLUSIONS Multiple use venting devices can be associated with cross contamination of blood cultures. This may result from contamination of the internal lumen of the venting device which is not exposed to the biocide. Medical microbiologists should consider the possibility of cross contamination associated with venting procedures when interpreting blood culture results. Further development of multiple use venting devices is required to reduce the risk of cross contamination of cultures.
منابع مشابه
آلودگی باکتریال کشتهای خون مرکز طبی کودکان تهران - فروردین تا خرداد 1383
Background: Blood culture is the criterion standard for identifying children with bacteremia. However, elevated false-positive rates are common and are associated with substantial health care costs. The aims of this prospective study were to: 1) determine the rate of blood culture contamination 2) determine variety and frequency of contaminant bacteria 3) compare the duration of hospital stay a...
متن کاملReduction in Blood Culture Contamination Through Use of Initial Specimen Diversion Device
Background Blood culture contamination is a clinically significant problem that results in patient harm and excess cost. Methods In a prospective, controlled trial at an academic center Emergency Department, a device that diverts and sequesters the initial 1.5-2 mL portion of blood (which presumably carries contaminating skin cells and microbes) was tested against standard phlebotomy procedur...
متن کاملBlood culture cross contamination associated with a radiometric analyzer.
During a 9-day period in August 1980 in a New Jersey hospital, three pairs of consecutively numbered blood cultures from different patients were identified as positive for the same organism (two pairs of Klebsiella pneumoniae and one pair of group A Streptococcus), for each pair, both cultures were positive in the same atmosphere, both organisms had the same sensitivities, and the second of eac...
متن کاملPlatelet Septic Transfusion Reactions in Patients With Hemato-Oncological Diseases
Background & Objective: Bacterial, contamination of blood components are a significant risk for transfusion reactions. Inherently, platelet concentrates (PCs) are vulnerable to bacterial contamination, due to the storage condition of processed PCs at room temperature, which provide very suitable conditions for the proliferation of microorganisms....
متن کاملMicrobial Profile of Air Contamination in Hospital Wards
Background and Aims:Nosocomial infections cause considerable morbidity and mortality and pose high financial burden on healthcare systems. Although surface contact, surgical incisions, wounds and catheters are responsible for a high percentage of nosocomial infections, bacterial and fungal air contaminations in hospitals have an important role in development of hospital infections. The p...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of clinical pathology
دوره 52 3 شماره
صفحات -
تاریخ انتشار 1999