Dynamics of cryogen deposition relative to heat extraction rate during cryogen spray cooling

نویسندگان

  • Wim Verkruysse
  • Boris Majaron
  • Guillermo Aguilar
  • Lars O Svaasand
  • Stuart Nelson
چکیده

Goal is to investigate how delivery nozzle design influences the cooling rate of cryogen spray as used in skin laser treatments. Cryogen was sprayed through nozzles that consist of metal tubes with either a narrow or wide diameter and two different lengths. Fast-flashlamp photography showed that the wide nozzles, in particular the long wide one, produced a cryogen jet (very small spray cone angle) rather than a spray (cone angles of about 15° or higher) and appeared to atomize the cryogen less finely than the narrow nozzles. We measured the cooling rate by spraying some cryogen on an epoxy -block with thermocouples embedded. The heat extraction rate of the wide nozzles was higher than that of the narrow nozzles. The results suggest that finely atomized droplets produced by the narrow nozzles do not have enough kinetic energy to break through a layer of liquid cryogen accumulated on the object, which may act as a thermal barrier and, thus, slow down heat extraction. Presumably, larger droplets or non-broken jets ensure a more violent impact on this layer and therefore ensure an enhanced thermal contact. The margin of error for the heat extraction estimate is analyzed when using the epoxy -block. We introduce a complementary method for estimating heat extraction rate of cryogen sprays.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Intermittent cryogen spray cooling for optimal heat extraction during dermatologic laser treatment.

Fast heat extraction is critically important to obtain the maximal benefit of cryogen spray cooling (CSC) during laser therapy of shallow skin lesions, such as port wine stain birthmarks. However, a film of liquid cryogen can build up on the skin surface, impairing heat transfer due to the relatively low thermal conductivity and higher temperature of the film as compared to the impinging spray ...

متن کامل

Influence of Cryogen Spray Cooling Parameters on the Heat Extraction Rate from a Sprayed Surface

Cryogen spray cooling is used to prevent epidermal thermal damage during port-wine stain laser therapy, despite the limited understanding of the fluid dynamics, thermodynamics, and heat transfer characteristics of cryogen sprays. In recent studies, it has been suggested that the heat flux through human skin could be increased by changing physical parameters such as nozzle-to-skin distance, nozz...

متن کامل

Cryogen spray cooling in laser dermatology: effects of ambient humidity and frost formation.

BACKGROUND AND OBJECTIVE Dynamics of cryogen spray deposition, water condensation and frost formation is studied in relationship to cooling rate and efficiency of cryogen spray cooling (CSC) in combination with laser dermatologic surgery. STUDY DESIGN/MATERIALS AND METHODS A high-speed video camera was used to image the surface of human skin during and after CSC using a commercial device. The...

متن کامل

Heat transfer between single cryogen droplet and epoxy skin phantom

Cryogen Spray Cooling (CSC) is an auxiliary procedure that pre-cools the epidermis during Laser Dermatologic Surgery (LDS) to avoid non-specific epidermal thermal damage. During CSC, explosive atomization, inflight evaporation and droplet-substrate interaction affect heat extraction from the substrate. In order to understand the heat transfer mechanism of this complex process, we first study th...

متن کامل

Heat-transfer dynamics during cryogen spray cooling of substrate at different initial temperatures.

Cryogen spray cooling (CSC) is used to minimize the risk of epidermal damage during laser dermatologic therapy. However, the dominant mechanisms of heat transfer during the transient cooling process are incompletely understood. The objective of this study is to elucidate the physics of CSC by measuring the effect of initial substrate temperature (T0) on cooling dynamics. Cryogen was delivered b...