Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population
نویسندگان
چکیده
Introduction The Sysmex XN-series haematology analyser has newly adopted a fluorescent channel to measure immature platelet fraction (IPF). To promote the clinical utility of this promising parameter, establishing a reliable reference interval is mandatory. According to previous studies, IPF values may be affected by the employed analyser and the ethnic background of the individual, but no differences seem to be found between individuals' genders. Therefore, this study aimed to define the reference interval for IPF in a Spanish population following Clinical and Laboratory Standard Institute (CLSI) guidelines. Materials and methods A total of 153 healthy Caucasian adults from Spain met the inclusion criteria. IPF measurement was performed by means of a Sysmex XN-2000 haematology analyser. A non-parametric percentile method was used to calculate the reference intervals in accordance with CLSI guidelines. Results The obtained reference interval for IPF on the Sysmex XN-2000 was 1.6-9.6% (90% confidence intervals (CIs) were 1.5-1.8 and 9.3-11.5, respectively). No significant gender difference in IPF reference intervals was observed (P = 0.101). Conclusions This study provides, for the first time, a reference interval for IPF using a Sysmex XN-2000 in a Spanish population, ranging from 1.6 to 9.6%. These data are needed to evaluate platelet production in several conditions such as thrombocytopenia, inflammatory states and cardiovascular diseases, as well as for future research.
منابع مشابه
Immature platelet fraction measured on the Sysmex XN hemocytometer predicts thrombopoietic recovery after autologous stem cell transplantation
OBJECTIVES A period of thrombocytopenia is common after stem cell transplantation (SCT). To prevent serious bleeding complications, prophylactic platelet transfusions are administered. Previous studies have shown that a rise in immature platelets precedes recovery of platelet count. Our aim was to define a cutoff value for immature platelets predicting thrombopoietic recovery within 2 d. METH...
متن کاملNew quantitative parameters on a recently introduced automated blood cell counter--the XE 2100.
The XE 2100 (Sysmex Corporation) is a cell counter that furthers the technology of fluorescent flow cytometry developed from the earlier range of Sysmex analysers. The new diagnostic features are a nucleated red cell count (NRBC), the ability to measure platelets by impedance as well as an 'optical' platelet count using a fluorescence dye and an immature granulocyte (IG) count. The NRBC count w...
متن کاملEstablishment of Age- and Gender-Specific Reference Ranges for 36 Routine and 57 Cell Population Data Items in a New Automated Blood Cell Analyzer, Sysmex XN-2000.
We established age- and gender-specific reference ranges for the 36 routine complete blood cell (CBC) and 57 cell population data (CPD) items in the Sysmex XN-2000 (Sysmex, Japan). In total, 280 peripheral blood samples were obtained from an equal number of healthy adults. Values for 36 routine items and 57 CPD items were obtained for each sample, and the results were categorized into six subgr...
متن کاملImmature Granulocyte Count on the New Sysmex Xn9000: Performance and Diagnosis of Sepsis in the Intensive Care Unit
Introduction. The amount of immature leukocytes reflects marrow response to bacterial infection, and this may be quantified as the band or immature granulocyte (IG) count. The aim of this study was to analyze the IG count performance of the Sysmex XN-9000 hematology analyzer in intensive care unit (ICU) patients. Methods. 480 peripheral blood samples from adult patients admitted to the ICU (301...
متن کاملAssessment of an immature platelet fraction (IPF) in peripheral thrombocytopenia.
A new automated method to reliably quantify reticulated platelets, expressed as the immature platelet fraction (IPF), has been developed utilizing the XE-2100 blood cell counter with upgraded software (Sysmex, Kobe, Japan). The IPF is identified by flow cytometry techniques and the use of a nucleic acid specific dye in the reticulocyte/optical platelet channel. The clinical utility of this para...
متن کامل