Anodal capture in cardiac resynchronization therapy implications for device programming.
نویسندگان
چکیده
INTRODUCTION Right ventricular (RV) anodal capture (AC) has been reported in cardiac resynchronization therapy (CRT), when left ventricular (LV) pacing uses pseudobipolar (LV tip to RV proximal electrode) configuration. The aim of the study was to analyze the prevalence of AC and its implications for device programming. METHODS AND RESULTS When AC occurred, the resulting QRS morphology was evaluated with the following pacing modes: (1) LV tip pacing plus RV AC, (2) Biventricular (BiV) pacing (i.e., both LV and RV tip pacing), and (3) BiV pacing plus RV AC. Several interventricular pacing (VV) intervals from 50 ms of LV preactivation to 30 ms of RV preactivation were tested in modes 2 and 3. From 38 consecutive patients, AC was achieved in 14 (in 74% of the pacemakers and in none of the defibrillators). LV tip pacing plus RV AC obtained narrower QRS than BiV pacing at all VV intervals in seven of the patients with AC (50%). When BiV pacing is combined with RV AC, it produced a ventricular depolarization through two wave fronts (one from the LV tip and the second from either the ring or the tip of the RV lead depending on the VV interval programmed). CONCLUSIONS AC obtained the narrowest QRS of all tested pacing modes in a significant proportion of patients undergoing CRT. Though the stimulus was delivered from three sites (BiV pacing plus RV AC mode), only two wave fronts of ventricular activation were seen by ECG.
منابع مشابه
Anodal Stimulation: An Underrecognized Cause of Nonresponders to Cardiac Resynchronization Therapy
OBJECTIVE The purpose of this study was to determine if anodal stimulation accounts for failure to benefit from cardiac resynchronization therapy (CRT) in some patients. BACKGROUND Approximately 30-40% of patients with moderate to severe heart failure do not have symptomatic nor echocardiographic improvement in cardiac function following CRT. Modern CRT devices allow the option of programming...
متن کاملOptimal programming in cardiac resynchronization therapy.
Cardiac resynchronization therapy (CRT) has been shown to improve cardiac function, diminish hospitalization frequency, and enhance quality of life in selected heart failure patients. This benefit is mostly due to improved synchronization of ventricular contraction in the diseased heart. Since heart failure patients represent a heterogeneous group, cardiac resynchronization therapy must be tail...
متن کاملUsefulness of the 12-lead electrocardiogram in the follow-up of patients with cardiac resynchronization devices. Part II.
The interval from the pacemaker stimulus to the onset of the earliest paced QRS complex (latency) may be prolonged during left ventricular (LV) pacing. Marked latency is more common with LV than right ventricular (RV) pacing because of indirect stimulation through a coronary vein and higher incidence of LV pathology including scars. During simultaneous biventricular (BiV) pacing a prolonged lat...
متن کاملIntentional anodal capture of a left ventricular quadripolar lead enhances resynchronization equally with multipoint pacing
Figure 1 Changes in the 12-lead electrocardiograms (ECGs) during biventricular pacing with different left ventricular (LV) pacing configurations. A: Extended bipolar pacing between LV1 and right ventricular coil. B: Extended bipolar pacing between LV4 and right ventricular coil. C: LV bipolar pacing between LV1 (anode) and LV4 (cathode) with an output of 4.0 V/0.4 ms. The similar morphology to ...
متن کاملSuccessful treatment of a cardiac resynchronization therapy nonresponder by identifying lead malpositioning.
This case describes some of the commonly overlooked device-related issues in patients who have reportedly failed to respond to cardiac resynchronization therapy (CRT). The case demonstrates voltage-dependent right ventricular capture instead of right atrial capture by a subtly malpositioned right atrial lead. CRT therapy failed to improve symptoms of heart failure and the diagnosis of "CRT nonr...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Pacing and clinical electrophysiology : PACE
دوره 29 9 شماره
صفحات -
تاریخ انتشار 2006