Controlled comparative evaluation of BacT/Alert FAN and ESP 80A aerobic media as means for detecting bacteremia and fungemia.

نویسندگان

  • G V Doern
  • A Barton
  • S Rao
چکیده

During a one-year period, a total of 6,305 blood cultures were processed in a tertiary-care teaching hospital; 6 to 12 ml of blood was inoculated into both a BacT/Alert Fan aerobic bottle and an ESP 80A aerobic bottle. The FAN aerobic bottle contains an antimicrobial-absorbing material; the 80A aerobic bottle does not. Bottles were processed on their respective continuous-monitoring blood culture instruments for up to five days of incubation. Four hundred thirty-three cultures (6.9%) representing 301 septic episodes in 235 different patients yielded 490 bacteria or yeasts thought to be clinically significant. Two hundred seventy-five of the 433 presumed clinically significant positive cultures (63.5%) representing 195 septic episodes and yielding 301 isolates were positive in both FAN and 80A bottles. One hundred nine significant positive cultures (25.2%) (i.e., cultures positive with an organism judged to be of probable clinical significance) from 70 septic episodes yielded 126 isolates only in FAN bottles. Conversely, the 80A bottle was exclusively positive in 49 instances (11.3%), representing 36 septic episodes and yielding 63 isolates. The higher rates of significant positive blood cultures, numbers of septic episodes documented, and numbers of isolates recovered in FAN bottles versus 80A bottles were all statistically significant (P < 0.05). Enhanced rates of detection of presumed clinically significant isolates in FAN bottles were largely accounted for by Staphylococcus aureus, members of the Enterobacteriaceae, and non-Pseudomonas aeruginosa miscellaneous gram-negative bacilli from patients receiving antimicrobial therapy at the time blood cultures were obtained. Enhanced recovery of one organism group, the beta-hemolytic streptococci, occurred in 80A. With one exception, detection times were essentially equivalent in the two systems. The single exception pertained to streptococci and enterococci, which were recovered significantly faster in 80A bottles. Three hundred thirty-eight of the 6,305 blood cultures evaluated in this study (5.4%) were judged likely to be contaminated. The percentages of probable contaminated cultures were as follows: 26.6% FAN and 80A; 42.3% FAN only; 31.1% 80A only (P < 0.05). Finally, the instrument false-positive rates for the two systems were 0.7% with FAN and 3.0% with 80A (P < 0.05). We conclude that while contamination rates were slightly higher with FAN than with 80A, use of FAN aerobic bottles in conjunction with the BacT/Alert system will yield significantly higher numbers of clinically significant blood culture isolates than 80A bottles and the ESP system. Furthermore, this enhanced detection is most conspicuous in patients receiving antimicrobial therapy at the time blood cultures are performed, probably due to the presence of an antimicrobial-absorbing material in FAN aerobic bottles.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Controlled comparison of BacT/ALERT FAN aerobic medium and BATEC fungal blood culture medium for detection of fungemia.

Yeasts are an increasingly common cause of nosocomial bloodstream infections. Methods for their detection are many; controlled comparisons are few. The vented FAN aerobic blood culture medium has been shown to be superior to the standard BacT/ALERT aerobic medium for the detection of fungemia as well as bacteremia. The BACTEC selective fungal medium (FM) (BD Biosciences, Sparks, Md.) allowed de...

متن کامل

Use of the BacT/Alert blood culture system for culture of sterile body fluids other than blood.

Studies have demonstrated that large-volume culture methods for sterile body fluids other than blood increase recovery compared to traditional plated-medium methods. BacT/Alert is a fully automated blood culture system for detecting bacteremia and fungemia. In this study, we compared culture in BacT/Alert standard aerobic and anaerobic bottles, BacT/Alert FAN aerobic and FAN anaerobic bottles, ...

متن کامل

Clinical comparison of nonvented aerobic BacT/Alert blood culture bottle and standard aerobic bottle for detection of microorganisms in blood.

The current BacT/Alert standard aerobic (VA) blood culture bottle was redesigned and designated a nonvented aerobic (NVA) culture bottle; this bottle does not require venting. A total of 3,873 sets of blood samples for culture were obtained from adult patients with suspected bacteremia or fungemia. The NVA bottle showed performance equivalent to that of the VA bottle for recovery and speed of d...

متن کامل

Controlled comparison of the BacT/Alert and BACTEC 660/730 nonradiometric blood culture systems.

In a collaborative study at three university hospitals, the recovery of microorganisms and the speed of detection of microbial growth by the BacT/Alert (Organon Teknika Corporation, Durham, N.C.) and BACTEC 660/730 (Becton-Dickinson Diagnostic Instrument Systems, Sparks, Md.) nonradiometric blood culture systems were compared. A total of 5,918 comparisons were made between BacT/Alert aerobic an...

متن کامل

Controlled clinical comparison of the BacT/ALERT FN and the standard anaerobic SN blood culture medium.

To determine the optimal anaerobic companion bottle to pair with the BacT/ALERT (bioMerieux, Durham, N.C.) nonvented aerobic FA (FA) medium for recovery of pathogenic microorganisms from adult patients with bacteremia and fungemia, we compared the BacT/ALERT FN (FN) anaerobic bottle with the standard BacT/ALERT SN (SN) anaerobic bottle. Each bottle, FA, FN, and SN, was filled with 8 to 12 ml of...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 36 9  شماره 

صفحات  -

تاریخ انتشار 1998