Behavior of frozen serum pools and lyophilized sera in an external quality-assessment scheme.
نویسندگان
چکیده
To further improve analytical accuracy in clinical chemistry, proficiency testing needs amelioration in the quality of materials tested and in target value assignment. To obtain information on the actual state-of-the-art in the Lombardy region of Italy, and to examine the behavior of different types of control materials (fresh-frozen human sera and lyophilized materials), we developed the following experimental design. Two human serum pools and two lyophilized sera were distributed to 32 laboratories for determination of glucose, creatinine, cholesterol, sodium, potassium, and gamma-glutamyltransferase (gamma-GT). Each analyte was measured in triplicate on each of 3 days. Target values for the controls were obtained with Reference Methods. The results show a good intralaboratory precision for every component but some accuracy problems for glucose, creatinine, cholesterol, and gamma-GT. Lyophilized materials showed some commutability problems for glucose, electrolytes, and gamma-GT, mainly with dry chemistry technology.
منابع مشابه
Creatinine measurement proficiency testing: assignment of matrix-adjusted ID GC-MS target values.
The results of an external quality-assessment experiment for serum creatinine measurement are described. Fifty-one laboratories performed quintuplicate analyses during three different analytical runs on six lyophilized sera and two frozen human serum pools. Isotope dilution gas chromatography-mass spectrometry (ID GC-MS) target values were assigned to all the materials. Intralaboratory within- ...
متن کاملTesting the accuracy of total cholesterol assays in an external quality-control program. Effect of adding sucrose to lyophilized control sera compared with use of fresh or frozen sera.
We studied the suitability of various types of human serum preparations to test the accuracy of total cholesterol measurements in the External Quality Assessment scheme in The Netherlands, in which approximately 180 laboratories participate. Checked against the certified Abell/Kendall Reference Method, large reagent-dependent negative biases were observed with lyophilized serum that was insuffi...
متن کاملStability of frozen liquid control sera for assay of cholesterol in high-density lipoprotein.
Liquid serum pools with low-, normal-, or above-normal concentration of high-density lipoprotein cholesterol were prepared by selection and dilution of sera having low lipoprotein content or by enrichment with concentrated high-density lipoprotein. Stability of the serum pools depended on storage conditions. On storage between -2 and -12 degrees C, the high-density lipoprotein cholesterol conte...
متن کاملEvaluation of quality-control materials for high-density-lipoprotein cholesterol quantitation.
Commercial quality-control materials and frozen plasma and serum pools were evaluated for high-density lipoprotein (HDL) cholesterol quantitation by the heparin-Mn2+ and phosphotungstate-Mg2+ precipitation methods and by ultracentrifugal and electrophoretic methods. We evaluated among-day precision; inter-method differences in HDL cholesterol quantitation; lipid composition and electrophoretic ...
متن کاملEffects of lyophilization of serum on the measurement of apolipoproteins A-I and B.
A common accuracy-based standardization program is indispensable for establishing reference intervals for the clinical use of apolipoproteins. The development and distribution of reference materials and quality-control materials that do not exhibit matrix effects between methods is essential to the standardization process. We examined the suitability of lyophilized material as a common referenc...