Measurement method of AC current up to 1 MHz
نویسندگان
چکیده
Wideband current transducers with a voltage output are often used to measure AC currents at frequencies up to 1 MHz. The ratio of output voltage to primary current as a function of frequency is the calibration result of these transducers. A traceable method to measure 1 A alternating current at frequencies of 100 kHz and 1 MHz is reported in this paper. The ratio of the wideband current transducer is obtained with expanded uncertainties of 1.3 % for 1 A, 100 kHz and of 2.0 % for 1 A, 1 MHz.
منابع مشابه
Spatial detection of ferromagnetic wires using GMR sensor and based on shape induced anisotropy
The purpose of this paper is to introduce a new technique for row spacing measurement in a wire array using giant magnetoresistive (GMR) sensor. The self-rectifying property of the GMR-based probes leads to accurately detection of the magnetic field fluctuations caused by surface-breaking cracks in conductive materials, shape-induced magnetic anisotropy, etc. The ability to manufacture probes h...
متن کاملCCE 92-05 intercomparison of ac-dc voltage transfer standards at high frequencies (1-50 MHz)
From August 1995 to May 1998, the CCE 92-05 intercomparison of ac–dc voltage transfer standards at high frequencies was carried out. Two travel standards were measured by 15 national standards institutes. The results in the frequency range from 1 to 50 MHz show a good agreement between the majority of participants. The span of the majority of the reported ac–dc differences at 50 MHz is less tha...
متن کاملError Analysis, Design and Modeling of an Improved Heterodyne Nano-Displacement Interferometer
A new heterodyne nano-displacement with error reduction is presented. The main errors affecting the displacement accuracy of the nano-displacement measurement system including intermodulation distortion error, cross-talk error, cross-polarization error and phase detection error are calculated. In the designed system, a He-Ne laser having three-longitudinal-mode is considered as the stabiliz...
متن کاملA Primary Level Ultrasonic Power Measurement System Developed at National Institute of Metrology, Thailand
The total output power from a medical ultrasound transducer has to be measured due to various reasons but in general they are related to assessment of performance and safety of equipment. The objective of this work was to develop a primary level ultrasonic power measurement system to establish standards for ultrasonic power measurement in Thailand. This system is being set up at the National In...
متن کاملThe sensitivity of a method to predict a capacitor's frequency characteristic
A joint effort between the U.S. Naval Academy and the National Institute of Standard and Technology (NIST) resulted in the development of a method to characterize the capacitance and dissipation factor of a set of commercial standard four terminal-pair (4TP) capacitors. The method depends on network analyzer impedance measurements at high frequencies (40 MH–200 MHz) and a regression of these me...
متن کامل