Calibration methods for a multi-channel multi-frequency EIT system.

نویسندگان

  • Tong In Oh
  • Kyung Heon Lee
  • Sang Min Kim
  • Hwan Koo
  • Eung Je Woo
  • David Holder
چکیده

Multi-channel multi-frequency electrical impedance tomography (EIT) systems require a careful calibration to minimize systematic errors. We describe novel calibration methods for the recently developed KHU Mark1 EIT system. Current source calibration includes maximization of output resistance and minimization of output capacitance using multiple generalized impedance converters. Phase and gain calibrations are used for voltmeters. Phase calibration nulls out the total system phase shift in measured voltage data. Gain calibrations are performed in two steps of intra- and inter-channel calibrations. Intra-channel calibration for each voltmeter compensates frequency dependence of its voltage gain and also discrepancy between design and actual gains. Inter-channel calibration compensates channel-dependent voltage gains of all voltmeters. Using the calibration methods described in this paper, we obtained 1 MOmega minimal output impedance of the current source in the frequency range 10 Hz-500 kHz. The reciprocity error was as small as 0.05% after intra- and inter-channel voltmeter calibrations. To demonstrate effects of calibration in reconstructed images, we used a homogenous phantom from which uniform images should be produced. Reconstructed time- and frequency-difference images using uncalibrated data showed spurious anomalies. By using calibrated data, standard deviations of time- and frequency-difference images of the homogenous phantom were reduced by about 40% and 90%, respectively.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Design and calibration of a compact multi-frequency EIT system for acute stroke imaging.

A new, compact UCLH Mk 2.5 EIT system has been developed and calibrated for EIT imaging of the head. Improvements include increased input and output impedances, increased bandwidth and improved CMRR (80 dB) and linearity over frequencies and load (0.2% on a single channel, +/-0.7% on a saline tank over 20 Hz-256 kHz and 10-65 Omega). The accuracy of the system is sufficient to image severe acut...

متن کامل

Electrical Impedance Tomography System for detecting fast physiological signal: KHU Mark2.5

Electrical impedance Tomography (EIT) can produce real-time images of conductivity variations associated with respiratory and cardiac cycles. We report the development of a new multi-frequency EIT system called the KHU Mark2.5. It has new features such as flexible electrode configurations to accommodate application-specific requirements, multiple independent current sources and voltmeters for f...

متن کامل

Time-Varying Frequency Fading Channel Tracking In OFDM-PLNC System, Using Kalman Filter

Physical-layer network coding (PLNC) has the ability to drastically improve the throughput of multi-source wireless communication systems. In this paper, we focus on the problem of channel tracking in a Decode-and-Forward (DF) OFDM PLNC system. We proposed a Kalman Filter-based algorithm for tracking the frequency/time fading channel in this system. Tracking of the channel is performed in the t...

متن کامل

Performance improvements in a MF-EIT system for acute stroke: The UCL Mk2.5

A recently developed EITS system, the UCLH mk2.5 was tested on a resistor phantom and in linear time and frequency difference imaging of a cylindrical tank, 10cm diameter, filled with 0.1% saline, and 16 electrodes in a ring; the test object was a cylinder of banana, 2 cm long and 1 cm in diameter. Reproducible images with the same location as the test object were obtained with both methods. No...

متن کامل

Comparison of Single- site and Multi-site Based Calibrations of SWAT in Taleghan Watershed, Iran

Calibration of model is critical for hydrologic modeling of large watersheds in a mountain watershed. In this study Soil and Water Assessment Tool (SWAT) used to comparison a single-site calibration procedure that employed streamflow measurement at outlet of watershed to a multi-site calibration method that used streamflow measurements at three stations (Galinak, Joestan and Dehdar). Results sh...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Physiological measurement

دوره 28 10  شماره 

صفحات  -

تاریخ انتشار 2007