Leukocyte counts in cerebrospinal fluid with the automated hematology analyzer CellDyn 3500 and the urine flow cytometer UF-100.
نویسندگان
چکیده
BACKGROUND The counting of leukocytes and erythrocytes in cerebrospinal fluid (CSF) is still performed microscopically, e.g., using a chamber in most laboratories. This requires sufficient practical experience, is time-consuming, and may constitute a problem in emergency diagnostics. Specific automated systems for CSF cell counting are not available at present. METHODS We tested the hematology analyzer CellDyn 3500 (CD) and the urine flow cytometer UF-100 (UF), which are not designed for CSF analysis. We studied >104 samples with both analyzers, and the counts obtained were compared with the reference method (Fuchs-Rosenthal chamber). RESULTS Good linearity in the medically relevant range of 15 x 10(6) to 1000 x 10(6) leukocytes/L and a high degree of within-run accuracy were seen for both analyzers. Cell counting on the UF was excellent, especially when low cell counts were encountered (CV, 4. 9% compared with 28% observed for the CD). Method comparison showed that identical results could be detected for a majority of the count pairs. For a few samples, there was a discrepancy between the results from the analyzers and the counting chamber. In most cases, these were CSF samples containing a high proportion of lymphocytes. For these samples, the CD result led to a false-positive high leukocyte count, and on the UF these cells were not allocated to the leukocyte population, thus leading to false-negative counts. CONCLUSIONS Both analyzers should not be used for CSF cell counting in all cases at present. However, once the technical and software problems have been solved, routine use of the two analyzers for CSF analysis should be seriously contemplated.
منابع مشابه
Automated flow cytometric analysis of cerebrospinal fluid.
BACKGROUND Recently, the UF-100 (Sysmex Corporation) flow cytometer was developed to automate urinalysis. We evaluated the use of flow cytometry in the analysis of cerebrospinal fluid (CSF). METHODS UF-100 data were correlated with microscopy and biochemical data for 256 CSF samples. Microbiological analysis was performed in 144 suspected cases of meningitis. RESULTS Good agreement was obta...
متن کاملAutomated flow cytometry compared with an automated dipstick reader for urinalysis.
Recently, the Sysmex UF-100 flow cytometer was developed to automate urinalysis. We compared UF-100 test results with those of an automated dipstick reader. A cross-check of UF-100, dipstick, and microscopic sediment data was performed in 1001 urine samples. Good agreements (P <0.001) were obtained between UF-100 and dipstick data for erythrocytes (r = 0.636) and leukocytes (r = 0.785). Even in...
متن کاملUnsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures.
UF-100 flow cytometer and urine strip results were cross-interpreted to predict culture outcomes. The best negative predictive value was obtained with bacteria at > or =1,000/microl, white blood cells at > or =20/microl, or leukocyte esterase positivity. Nine of 24 false negatives were clinically significant. Thus, UF-100 and urine strip results do not accurately predict the outcome of cultures.
متن کاملEvaluation of the Sysmex UF-100 urine cell analyzer as a screening test to reduce the need for urine cultures for community-acquired urinary tract infection.
We evaluated the UF-100 flow cytometer (TOA Medical Electronics, Kobe, Japan) as a screening test for community-acquired urinary tract infection (UTI) to reduce the need for bacterial cultures. By comparing the test results for 330 urine samples with quantitative urine cultures, we established cutoff criteria for the UF-100. To rule out hospital-acquired UTI, all urine samples were from new pat...
متن کاملAutomated urinalysis: first experiences and a comparison between the Iris iQ200 urine microscopy system, the Sysmex UF-100 flow cytometer and manual microscopic particle counting.
BACKGROUND Automated analysis of insoluble urine components can reduce the workload of conventional microscopic examination of urine sediment and is possibly helpful for standardization. We compared the diagnostic performance of two automated urine sediment analyzers and combined dipstick/automated urine analysis with that of the traditional dipstick/microscopy algorithm. METHODS A total of 3...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 46 2 شماره
صفحات -
تاریخ انتشار 2000