Automated measurement of 25-OH vitamin D3 on the Roche Modular E170 analyzer.

نویسندگان

  • Aila Leino
  • Ursula Turpeinen
  • Pertti Koskinen
چکیده

BACKGROUND The first commercial direct automated immunoassay specific for 25-OH vitamin D(3) (25-OH-D(3)) was recently introduced for use on Roche Diagnostics immunoassay analyzers. We assessed the analytical performance of the Elecsys 25-OH-D(3) assay on a Roche Modular E 170 analyzer. METHODS The Elecsys 25-OH-D(3) assay is a direct electrochemiluminescence immunoassay for human serum or plasma. It is a competitive assay in which the binding protein of vitamin D is inactivated during incubation. The assay employs a polyclonal antibody directed against 25-OH vitamin D(3). We compared the 25-OH-D(3) assay to assays performed with RIA, HPLC, and liquid chromatography-tandem mass spectrometry (LC-MS/MS). RESULTS At concentrations of 48, 76, and 124 nmol/L, within-run CVs were 5.1%, 3.1%, and 7.1% and total CVs were 12.1%, 7.4%, and 10.6%, respectively. A comparison of Elecsys 25-OH vitamin D(3) with RIA yielded the regression equation: Elecsys = 1.114 x RIA - 6.15 (S(y|x) = 15.7 nmol/L; n = 163). The corresponding equation with HPLC was: Elecsys = 1.077 x HPLC + 5.442 (S(y|x) = 13.9 nmol/L; n = 67) and with LC-MS/MS: Elecsys = 0.887 x LC-MS/MS + 5.046 (S(y|x) =12.4 nmol/L; n = 64). Contrary to LC-MS/MS, with the cutoff of 50 nmol/L (deficiency vs normal), approximately 10% of samples were misclassified as normal with RIA and Elecsys. Plasma samples were observed to have markedly higher concentrations than serum samples. CONCLUSIONS The Elecsys concentrations of 25-OH-D(3) were in good overall agreement with those determined with LC-MS/MS and RIA. However, large between-method variation was observed in individual patient samples. Use of serum rather than plasma is preferred owing to the higher results observed with plasma samples.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Performance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics E170 Total 25-OH Vitamin D assays.

OBJECTIVES To evaluate the newly developed Roche MODULAR Analytics E170 Total Vitamin D and the Siemens ADVIA Centaur Vitamin D Total assays. MATERIALS AND METHODS Assays were evaluated using the Clinical and Laboratory Standards Institute protocols. Split patient samples were compared with LC-MS/MS and DiaSorin LIAISON assays (n=79 including 15 specimens with detectable endogenous 25-OH vita...

متن کامل

Performance characteristics of six automated 25-hydroxyvitamin D assays: Mind your 3s and 2s.

BACKGROUND Measurement of 25(OH)D has evolved rapidly, with an increased number of high-throughput automated immunoassays becoming available in recent years. The aim of this study was to fully evaluate six commercially available automated 25(OH)D immunoassays, including 2 newly available (Beckman Coulter) and one recalibrated (Siemens) assay. Comparisons were made to specifically identify the e...

متن کامل

Evaluation of new Beckman Coulter 25(OH) Vitamin D assay and potential improvement of clinical interpretation

INTRODUCTION The aim of this study was to assess the analytical performances of the newly developed Access2 25-hydroxyvitamin D (25(OH)D) total immunoassay on two analysers, DxI800 and Access2 (Beckman Coulter, Brea, CA, USA), and compare these two and a recalibrated Modular E 170 25(OH)D assay (Roche Diagnostics, Penzberg, Germany) with reference liquid chromatography tandem-mass spectrometry ...

متن کامل

Total 25-OH Vitamin D Concentrations in Chinese, Malays and Indians

There has been increasing interest in recent years in the prevalence of low 25-hydroxy (OH) vitamin D concentrations in different human populations [1, 2]. There are 2 forms of vitamin D: cholecalciferol (vitamin D3) and ergocalciferol (vitamin D2). Vitamin D3 is manufactured through the photoisomerization of 7-dehydrocholesterol to pre-vitamin D3, which then undergoes a temperature-dependent r...

متن کامل

Serum 25-hydroxyvitamin D immunoassays: recommendations for correct clinical interpretation.

assay on the Roche Modular analyzer. They used clinical samples that were almost free of 25hydroxyvitamin D2 (25-OH-D2) ( 10 nmol/L) as confirmed by liquid chromatography–tandem mass spectrometry (LC-MS/MS) and found a good overall agreement between the results obtained with the Roche assay and those measured with the DiaSorin RIA and by LC-MS/MS. Although they acknowledged that the Roche assay...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 54 12  شماره 

صفحات  -

تاریخ انتشار 2008