Evaluation of an engineering model of the "EKTACHEM" analyzer for glucose and urea assay.

نویسندگان

  • J C Cate
  • R Hedrick
  • M Taylor
  • C D McGlothlin
چکیده

We report results of a 21-week evaluation of an engineering model of the Kodak “EKTACHEM Glu/BUN Analyzer.” This novel clinical instrument measures glucose and “urea nitrogen” (actually, urea) in a minimum of 30 j.tL of body fluids by use of the principles of thin, multilayer film analysis and reflectance spectrophotometry. We compared the performance of the instrument to that of the Dupont aca and the Beckman Glucose Analyzer for glucose, and to the Dupont aca and the Technicon SMA 12/60 for urea. We also measured glucose in cerebrospinal fluid, with use of aqueous standards. Instrument stability during 39 days without recalibration was evaluated. Results from the EKTACHEM compared favorably with those from the other instruments with respect to precision, accuracy, analytical recovery, instrument and reagent stability, and (non-drug) interferences. General instrument performance was excellent, with little downtime. The instrument was enthusiastically accepted by our laboratory personnel.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Extended clinical trial and evaluation of urea nitrogen determination with the Ektachem GLU/BUN analyzer.

We followed the "abbreviated precision protocol" of the National Committee for Clinical Laboratory Standards for the evaluation of precision, accuracy, and carryover in analysis for urea nitrogen with the multilayer film analysis system ("Ektachem"). We analyzed 456 clinical samples with this instrument, by the manual urease/glucose dehydrogenase method, and with the Beckman System I GLU/BUN An...

متن کامل

The determination of calcium, glucose, urea and uric acid using the Kodak EKTACHEM multilayer film technology: an evaluation.

Calcium, glucose, urea and uric acid determinations on the "'Kodak Ektachem Four Chemistry Analyzer" were tested for precision and an estimation of the accuracy was performed with our laboratory routine methods using patients' sera and proficiency fluids. Precision -- examined over a period of six weeks -- was very good. Even the very stringent claim of the College of American Pathologists (CAP...

متن کامل

Extended clinical trial and evaluation of glucose determination with the Eastman Kodak Ektachem GLU/BUN Analyzer.

We followed the "abbreviated precision protocol" of the National Committee for Clinical Laboratory Standards for the evaluation of precision, accuracy, and carryover in analyses for glucose with the "Ektachem." We analyzed 760 clinical samples by this technique, by the FDA Proposed Class Standard glucose reference method, and with the Beckman System I GLU/BUN Analyzer. Precision and accuracy we...

متن کامل

Abstracts/Sommaires/Zusammenfassungen

An evaluation of the Kodak glucose/BUN analyzer including experience with proposed testing protocols An evaluation of the analysis of serum urea and plasma_glucose on the Kodak Ektachem GLU/BUN()analyzer is described. It incorporates the experimental design and statistical methods outlined in the Proposed Standard for Establishing Performance Claims (PSEP) (1) for clinical chemical methods PSEP...

متن کامل

Applicability of the Ektachem 400 analyzer for assaying analytes in miscellaneous body fluids.

The Ektachem analyzer, a slide-film chemistry technology, demonstrated applicability to non-serum, non-urine stimulated and actual clinical specimens having a wide range of protein concentration and pH. The simulated specimens studied had pH and protein that ranged from 2 to 9 and 0 to 88 g/L, respectively. Actual patients' specimens studied, which had protein concentrations ranging from 0.1 to...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 26 2  شماره 

صفحات  -

تاریخ انتشار 1980