A prospective randomized equivalence trial of the GlideScope Cobalt® video laryngoscope to traditional direct laryngoscopy in neonates and infants.

نویسندگان

  • John E Fiadjoe
  • Harshad Gurnaney
  • Nicholas Dalesio
  • Emily Sussman
  • Huaqing Zhao
  • Xuemei Zhang
  • Paul A Stricker
چکیده

BACKGROUND Intubation in children is increasingly performed using video laryngoscopes. Many pediatric studies examine novice laryngoscopists or describe single patient experiences. This prospective randomized nonblinded equivalence trial compares intubation time for the GlideScope Cobalt® video laryngoscope (GCV, Verathon Medical, Bothell, WA) with direct laryngoscopy with a Miller blade (DL, Heine, Dover, NH) in anatomically normal neonates and infants. The primary hypothesis was that intubation times with GCV would be noninferior to DL. METHODS Sixty subjects presenting for elective surgery were randomly assigned to intubation using GCV or DL. Intubation time, time to best view, percentage of glottic opening score, and intubation success were documented. We defined an intubation time difference of less than 10 s as clinically insignificant. RESULTS There was no difference in intubation time between the groups (GCV median = 22.6 s; DL median = 21.4 s; P = 0.24). The 95% one-sided CI for mean difference between the groups was less than 8.3 s. GCV yielded faster time to best view (median = 8.1 s; DL 9.9 s; P = 0.03). Endotracheal tube passage time was longer for GCV (median = 14.3 s; DL 8.5 s; P = 0.007). The percentage of glottic opening score was improved with GCV (median 100; DL 80; P < 0.0001). CONCLUSIONS Similar intubation times and success rates were achieved in anatomically normal neonates and infants with the GCV as with DL. The GCV yielded faster time to best view and better views but longer tube passage times than DL.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Tracheal intubation performed with GlideScope® video laryngoscope and direct laryngoscopy in neonates and infants.

1. Aziz MF, Dillman D, Fu R, Brambrink AM: Comparative effectiveness of the C-MAC video laryngoscope versus direct laryngoscopy in the setting of the predicted difficult airway. ANESTHESIOLOGY 2012; 116:629 –36 2. Aziz MF, Healy D, Kheterpal S, Fu RF, Dillman D, Brambrink AM: Routine clinical practice effectiveness of the Glidescope in difficult airway management: An analysis of 2,004 Glidescop...

متن کامل

Comparison of hemodynamic changes in patients using conventional laryngoscopy and video laryngoscope in surgery

Introduction: One of the complications of laryngoscopy and endotracheal intubation is sympathetic stimulation with hemodynamic changes that are always tried to be avoided by anesthesiologists. The aim of this study is to review and compare hemodynamic changes caused by both conventional and video laryngoscopy. Methods: In this clinical trial performed in 2015 in the Birjand-based Imam Reza Hosp...

متن کامل

Comparison of video laryngoscopy versus direct laryngoscopy during urgent endotracheal intubation: a randomized controlled trial.

OBJECTIVES In the critically ill undergoing urgent endotracheal intubation by direct laryngoscopy, multiple attempts are often required with a higher complication rate due to the urgency, uncontrolled setting, comorbidities, and variability in expertise of operators. We hypothesized that Glidescope video laryngoscopy would be superior to direct laryngoscopy during urgent endotracheal intubation...

متن کامل

Comparison of the Incidence of Sore Throat and Hemodynamic Changes in ‎Airway Management by Macintosh, Glideoscope Laryngoscope and ‎Laryngeal Mask Airway Techniques: Randomized Controlled Trial

Background: Studies have shown that postoperative sore throat is the most common complication of tracheal intubation, However, the actual incidence and extent of it are not well documented. Objective: The aim of this study was to evaluate and compare the incidence of sore throat and hemodynamic Changes in Macintosh, Glidescope Laryngoscope and laryngeal mask airway  techniques in elective surg...

متن کامل

A Randomized, Blinded, Clinical Study of Injury Incidence During Endotracheal Intubation: Comparison of GlideScope Video Laryngoscopy and direct Laryngoscopy, AANA Journal, December 2017

AANA Journal  December 2017  Vol. 85, No. 6 445 Injuries can occur during endotracheal intubation. This prospective, randomized, blinded study aimed to determine if a difference exists in incidence of injury following endotracheal intubation using either direct laryngoscopy or the GlideScope video laryngoscope (Verathon Inc). A convenience sample of 155 patients scheduled to undergo surgery t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Anesthesiology

دوره 116 3  شماره 

صفحات  -

تاریخ انتشار 2012