Inadvertent Lead Placement In The Left Ventricle: A Case Report And Brief Review

نویسندگان

  • David D McManus
  • Mary-Lee Mattei
  • Karen Rose
  • Jason Rashkin
  • Lawrence S Rosenthal
چکیده

Inadvertent lead placement in the left ventricle (LV) is an uncommon and often under-diagnosed complication of cardiac device implantation. Thromboembolic (TE) events are common and usually secondary to fibrosis or thrombus formation on or around the lead. Anticoagulation can prevent TE events. Percutaneous and surgical LV lead extractions have been performed successfully, but the risks of percutaneous lead removal are not well-defined. In this report, we describe a case of inadvertent LV lead placement and briefly review the contemporary literature.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Conservative management of superior vena cava injury due to hemodialysis catheter placement: A case report and literature review

We reported the case of a 78-year old woman with the inadvertent superior vena cava injury resulting from hemodialysis catheter placement. The catheter was placed through the left internal jugular vein. Hemodynamics were stable. The initial chest film illustrated an extraluminal catheter and mediastinal widening. The patient did not have hemothorax. Computed tomography venography also confirmed...

متن کامل

Inadvertent malposition of a permanent pacemaker ventricular lead into the left ventricle which was initially missed and diagnosed two years later: a case report

INTRODUCTION Inadvertent malposition of a pacemaker ventricular lead into the left ventricle is an uncommon event, and its actual incidence is probably unknown. It may be underestimated and underreported because of a possible asymptomatic course. A 12-lead electrocardiogram is important to confirm proper placement. CASE PRESENTATION We report a case of a 60-year-old Caucasian man with a malpo...

متن کامل

Removal of Misplaced Left Ventricular Single Lead Pacemaker in a Patient Presenting with Recurrent Transient Ischemic Attacks

HINOJOS A, ILG K. Removal of Misplaced Left Ventricular Single Lead Pacemaker in a Patient Presenting with Recurrent Transient Ischemic Attacks. Spartan Med. Res. J. Vol. 2, No. 1, pp. 64-73, 2017. CONTEXT: Over 200,000 cardiac electronic implantable devices are annually placed in individuals living within the United States. Complications from this procedure can range up to 12%. Inadvertent lea...

متن کامل

Pacemaker lead malpositioning led to subsequent ischemic strokes despite antiplatelet and anticoagulation therapy

Pacemaker lead malpositioning may lead to severe clinical adverse events. Rarely, cases of inadvertent placement of a lead into the left ventricle are reported in the literature. We herein report a case of pacemaker lead malpositioning into the left ventricle via a persistent foramen ovale in a male caucasian patient. After this procedural adverse event, the patient suffered from two ischemic s...

متن کامل

Inadvertent placement of pacemaker lead into the systemic ventricle in repaired D-transposition of the great arteries.

We report a case of a 27-year-old man with Senning repair of D-transposition of the great arteries. A dual-chamber pacemaker was placed for syncope due to sinoatrial node dysfunction 3 years before this presentation. He had no further syncope but complained of dyspnea on exertion and mild exercise intolerance. A routine chest x-ray and transthoracic echocardiography suggested the presence of a ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 9  شماره 

صفحات  -

تاریخ انتشار 2009