Distal access using hyperflexible atraumatic distal tip with optimized proximal stability of the Benchmark intracranial guide catheter for the treatment of cerebral vascular diseases: a technical note
نویسندگان
چکیده
BACKGROUND A stable guide catheter position within the intracranial vasculature is critical for safe, successful endovascular treatment. OBJECTIVE To present ourinitial experience with the 0.071 inch inner diameter Benchmark guide catheter used in the treatment of intracranial cerebrovascular pathologies, demonstrating its safety and efficacy. METHODS We retrospectively reviewed use of the Benchmark guide catheter from September through December 2014 in the management of various neuroendovascular intracranial pathologies. Clinical performance and complication rates were evaluated, with particular consideration of vessel tortuosity. A total of 62 Benchmarks were used, 47 in the anterior circulation, 10 in the posterior circulation, 4 in the external carotid, and 1 in the venous sinus. The five cases with access to the external carotid and venous sinus were excluded. RESULTS The Benchmark was able to cross at least one 90° turn in 49 (86%) of the 57 patients. Reversal of the catheter was seen in 15% of 47 anterior circulation cases (4 at one 90° turn; 3 at two 90° turns). We report no complications of dissection or thromboembolic events. All guide catheter positions were safely achieved over a 0.035 Terumo stiff glidewire without need for an inner smaller lumen guide catheter for navigation. CONCLUSIONS Benchmark is a new guide catheter, with an ideal combination of both hyperflexible, atraumatic distal tip and optimized proximal shaft support to provide stable 6F primary access for a successful neurointerventional procedure. Benchmark can be easily, safely, and consistently positioned in a desired location within intracranial arteries providing a stable position for intervention and adequate angiography.
منابع مشابه
Z-Shaped Microcatheter Tip Shaping for Embolization of Aneurysms at the Proximal A1 Segment of the Anterior Cerebral Artery: A Technical Note
Embolization of an aneurysm at the proximal A1 segment of the anterior cerebral artery (ACA) is technically challenging. We treated four consecutive patients with unruptured aneurysms at the proximal portion of the A1 segment of the ACA. We tested various microcatheter tip shapes before we successfully embolized a lesion with a zigzag-shaped microcatheter tip, which was made by steam shaping. T...
متن کاملTransradial Neurointerventional Approach to Basilar Tip Aneurysm: Case Report and Technical Note
Cerebral angiography through transfemoral approach is preferred in most of the situation though in cardiovascular surgery transradial angiography is accounted as the superior standard. The present study aimed at presenting our early experience of transradial approach to cerebral angiography and intervention of a patient with basilar tip aneurysm that had inappropriate vasculature to perform rou...
متن کاملUtilization of a Novel, Multi-Durometer Intracranial Distal Access Catheter: Nuances and Experience in 110 Consecutive Cases of Aneurysm Flow Diversion.
BACKGROUND Coaxial catheter support systems provide a safe and stable foundation in endovascular treatment of intracranial aneurysms. Increasingly, robust distal intracranial support is sought during complex neurointerventions. The AXS Catalyst 5 distal access catheter (Cat5) is a new intracranial catheter designed for improved trackability and stability. We report the first experience using Ca...
متن کاملRisk of distal embolization with stent retriever thrombectomy and ADAPT.
BACKGROUND There is a discrepancy in clinical outcomes and the achieved recanalization rates with stent retrievers in the endovascular treatment of ischemic stroke. It is our hypothesis that procedural release of embolic particulate may be one contributor to poor outcomes and is a modifiable risk. The goal of this study is to assess various treatment strategies that reduce the risk of distal em...
متن کاملStructure of the orifice of the renal artery in the abdominal aorta in adult male dog
One of the locations of renal artery atherosclerosis is at the orifice of the renal artery, therefore thestructure of this orifice was assessed in 6 normal, adult male dogs by light microscopy (LM) andtransmission electron microscopy (TEM). For the LM study, processed tissues were embedded in paraffinand sectioned serially into 6-μm thickness. Sections were stained with orcein. For the TEM stud...