Direction and rate of socket wear in Charnley low-friction arthroplasty.
نویسنده
چکیده
Twenty-two high-density polyethylene sockets from Charnley low-friction arthroplasties have been studied. Acrylic casts and shadowgraph techniques were used to measure both the real and radiographic rates of wear; these rates showed a significant correlation. In the sagittal plane, nine of the sockets had worn lateral to a line drawn vertically from the centre of curvature of the socket, 12 had worn medial to that line and only one was worn exactly in the line. In the coronal plane, nine sockets had worn in front of a similar vertical line, two behind that line and 11 had worn exactly in the line. Evidence of impingement of the neck of the stem onto the rim of the socket was found in 14 patients; this is considered to be one of the causes of socket loosening. The obvious solution is to reduce the diameter of the neck of the stem from 12.5 mm to 10 mm; provided that it is made of cold-formed, high nitrogen-content stainless steel, this narrower neck is strong enough not to fracture.
منابع مشابه
Polyethylene wear in the Charnley offset bore acetabular cup. A radiological analysis.
We studied wear in the ultra-high-molecular-weight polyethylene offset bore socket in 54 hips which had had Charnley low-friction arthroplasty. At an average follow-up of 8.1 years, the mean penetration rate was 0.04 mm per year. Correlation between the depth of socket penetration and the incidence of socket migration was confirmed, but socket migration occurred with lower penetration than had ...
متن کاملOsteolysis after Charnley primary low-friction arthroplasty. A comparison of two matched paired groups.
We reviewed 249 consecutive Charnley primary low-friction arthroplasties in 191 patients performed by one surgeon using a transtrochanteric approach at a minimum follow-up of ten years. Of these, 37 hips in 32 patients showed osteolysis and were compared with 41 hips in 37 matched patients with no osteolysis. We assessed in each case the wear rate, stability of the prosthesis, acetabular angle,...
متن کاملRevision of the Charnley low-friction arthroplasty for recurrent or irreducible dislocation.
Twenty-one cases of the Charnley low-friction arthroplasty were revised because of recurrent or irreducible dislocation. Three main reasons for this revision, usually in combination, were identified: loss of the abductor mechanism due to trochanteric detachment; shortening of the limb due to high placement of the socket or low section of the femoral neck; and malorientation of the components. S...
متن کاملLong-term behaviour of the Charnley offset-bore acetabular cup.
We report the long-term radiological results of 58 total hip arthroplasties (THA) using the Charnley offset-bore acetabular socket. Wear was measured at four sites and radiolucent lines and possible migration were recorded. Four cups were retrieved at revision and were examined using light microscopy, SEM and X-ray microanalysis. At a mean follow-up of seven years the mean wear in the DeLee and...
متن کاملCharnley low-friction arthroplasty with an autograft of the femoral head for developmental dysplasia of the hip. The 10- to 15-year results.
Between 1983 and 1988 we carried out 45 Charnley low-friction arthroplasties with autografts from the femoral head in 41 patients for developmental dysplasia of the hip. The preoperative radiographs were assessed for the severity of DDH according to the classifications of Crowe et al, Hartofilakidis et al and Sharp. The postoperative and follow-up radiographs were analysed for coverage of the s...