The Heater Cooler as a Source of Infection from Nontuberculous Mycobacteria.
نویسندگان
چکیده
Nosocomial infections acquired during the course of cardiac surgery and hospitalization can have devastating patient consequences. The source of these infections is often difficult to determine which complicates eradication efforts. Recently it has become apparent that the heater-cooler devices used in conjunction with cardiopulmonary bypass may become contaminated with bacteria that are normally found in hospital water sources. The culprit organisms are nontuberculous mycobacteria which coat the intrinsic surfaces found within the circuits of the heater-coolers. Aerosolization of the bacteria occurs during normal heater-cooler operation which can disperse the organisms throughout the operating room. The bacteria are slow-growing and may not present for months, or years, following exposure which makes epidemiological determination a challenge. The ensuing report summarizes a recent outbreak in these infections that have been reported both in Europe and the United States, along with efforts to reduce the risk for patient infection.
منابع مشابه
Invasive Nontuberculous Mycobacterial Infections among Cardiothoracic Surgical Patients Exposed to Heater–Cooler Devices1
Invasive nontuberculous mycobacteria (NTM) infections may result from a previously unrecognized source of transmission, heater-cooler devices (HCDs) used during cardiac surgery. In July 2015, the Pennsylvania Department of Health notified the Centers for Disease Control and Prevention (CDC) about a cluster of NTM infections among cardiothoracic surgical patients at 1 hospital. We conducted a ca...
متن کاملReemergence of Mycobacterium chimaera in Heater–Cooler Units despite Intensified Cleaning and Disinfection Protocol
Invasive Mycobacterium chimaera infections after open-heart surgery have been reported internationally. These devastating infections result from aerosols generated by contaminated heater-cooler units used with extracorporeal circulation during surgery. Despite intensified cleaning and disinfection, surveillance samples from factory-new units acquired during 2014 grew nontuberculous mycobacteria...
متن کاملThe role of nontuberculous mycobacteria in cutaneous infections of diabetic patients: a literature review
Background: Diabetes is an important metabolic disease with myriad manifestations and complications, affecting many people. Cutaneous infections impact numerous diabetic patients. Among different bacterial infections in diabetic patients, the infections caused by mycobacteria other than Mycobacterium tuberculosis (MOTT) eventuate in complications owing to the paucity of accurate detection...
متن کاملOptimized Method for Isolation of Nontuberculous Mycobacteria from Hospital Aquatic Sources
Background: Aquatic ecosystems are an important source of nontuberculous mycobacteria (NTM) that can cause different diseases in human. Since culture of mycobacteria needs long-term incubation, fast-growing microorganisms and contaminants in the environment usually prevents the isolation of mycobacteria. Here, we compare different treatment protocols and describe a method that increases the rec...
متن کاملNotes from the Field: Mycobacterium chimaera Contamination of Heater-Cooler Devices Used in Cardiac Surgery - United States.
In the spring of 2015, investigators in Switzerland reported a cluster of six patients with invasive infection with Mycobacterium chimaera, a species of nontuberculous mycobacterium ubiquitous in soil and water. The infected patients had undergone open-heart surgery that used contaminated heater-cooler devices during extracorporeal circulation (1). In July 2015, a Pennsylvania hospital also ide...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The journal of extra-corporeal technology
دوره 48 2 شماره
صفحات -
تاریخ انتشار 2016