Tooth enamel EPR dosimetry: optimization of EPR spectra recording parameters and effect of sample mass on spectral sensitivity.
نویسندگان
چکیده
In order to improve the accuracy of the tooth enamel EPR dosimetry method, EPR spectra recording conditions were optimized. The uncertainty of dose determination was obtained as the mean square deviation of doses, determined with the use of a spectra deconvolution program, from the nominal doses for ten enamel samples irradiated in the range from 0 to 500 mGy. The spectra were recorded at different microwave powers and accumulation times. It was shown that minimal uncertainty is achieved at the microwave power of about 2 mW for a used spectrometer JEOL JES-FA100. It was found that a limit of the accumulation time exists beyond which uncertainty reduction is ineffective. At an established total time of measurement, reduced uncertainty is obtained by averaging the experimental doses determined from recorded spectra following intermittent sample shaking and sample tube rotation, rather than from one spectrum recorded at longer accumulation time. The effect of sample mass on the spectrometer's sensitivity was investigated in order to find out how to make appropriate corrections.
منابع مشابه
Human EPR Dosimetry at Low Accumulated Dose of Ionizing Radiation Human EPR Dosimetry at Low Accumulated Dose of Ionizing Radiation
A new method for tooth (enamel and dentine) and bone EPR dosimetry is proposed that enhances substantially the dosimetric sensitivity. The method is based on the nonstandard spectra registration mode (the second harmonic phase quadrature detection of absorption under rapid passage conditions). This method provides an apparent enhancement of the component ratio of radiation to a nonradiative bac...
متن کاملThe 4th international comparison on EPR dosimetry with tooth enamel Part 1: Report on the results
This paper presents the results of the 4th International Comparison of in vitro electron paramagnetic resonance dosimetry with tooth enamel, where the performance parameters of tooth enamel dosimetry methods were compared among sixteen laboratories from all over the world. The participating laboratories were asked to determine a calibration curve with a set of tooth enamel powder samples provid...
متن کاملDemystifying EPR: A Rookie Guide to the Application of Electron Paramagnetic Resonance Spectroscopy on Biomolecules
Electron Paramagnetic Resonance (EPR) spectroscopy, also known as Electron Spin Resonance(ESR) especially among physicists, is a strong and versatile spectroscopic method forinvestigation of paramagnetic systems, i.e. systems like free radicals and most transition metalions, which have unpaired electrons. The sensitivity and selectivity of EPR are notable andintriguing as compared to other spec...
متن کاملInvestigation of some parameters influencing the sensitivity of human tooth enamel to gamma radiation using electron paramagnetic resonance.
Electron paramagnetic resonance (EPR) has been successfully used as a physical technique for gamma radiation dose reconstruction using calcified tissues. To minimize potential discrepancies between EPR readings in future studies, the effects of cavity response factor, tooth position and donor gender on the estimated gamma radiation dose were studied. It was found that the EPR response per sampl...
متن کاملRetrospective dosimetry using fingernail in nuclear industries: A review paper
Generally, doses greater than 2Gy can produce acute biological effects in humans. For this reason, to measure and record the cumulative dose, especially in radiation users, the need for a personal dosimeter has been strongly felt by the researchers. The purpose of this study is to introduce retrospective dosimetric techniques, in particular the use of nails as a natural tool for determining cum...