Characterization of an orthovoltage biological irradiator used for radiobiological research
نویسندگان
چکیده
Orthovoltage irradiators are routinely used to irradiate specimens and small animals in biological research. There are several reports on the characteristics of these units for small field irradiations. However, there is limited knowledge about use of these units for large fields, which are essential for emerging large-field irregular shape irradiations, namely total marrow irradiation used as a conditioning regimen for hematological malignancies. This work describes characterization of a self-contained Orthovoltage biological irradiator for large fields using measurements and Monte Carlo simulations that could be used to compute the dose for in vivo or in vitro studies for large-field irradiation using this or a similar unit. Percentage depth dose, profiles, scatter factors, and half-value layers were measured and analyzed. A Monte Carlo model of the unit was created and used to generate depth dose and profiles, as well as scatter factors. An ion chamber array was also used for profile measurements of flatness and symmetry. The output was determined according to AAPM Task Group 61 guidelines. The depth dose measurements compare well with published data for similar beams. The Monte Carlo-generated depth dose and profiles match our measured doses to within 2%. Scatter factor measurements indicate gradual variation of these factors with field size. Dose rate measured by placing the ion chamber atop the unit's steel plate or solid water indicate enhanced readings of 5 to 28% compared with those measured in air. The stability of output over a 5-year period is within 2% of the 5-year average.
منابع مشابه
Biological X-ray irradiator characterization for use with small animals and cells
This study presents the characterization of an X-ray irradiator through dosimetric tests, which confirms the actual dose rate that small animals and cells will be exposed to during radiobiological experiments. We evaluated the linearity, consistency, repeatability, and dose distribution in the positions in which the animals or cells are placed during irradiation. In addition, we evaluated the p...
متن کاملCancer Cell Radiobiological Studies Using In-House-Developed α-Particle Irradiator.
An α-particle irradiator, enabling high-precision irradiation of cells for in vitro studies, has been constructed. The irradiation source was a (241)Am source, on which well inserts containing cancer cells growing in monolayer were placed. The total radioactivity, uniformity, and α-particle spectrum were determined by use of HPGe detector, Gafchromic dosimetry film, and PIPS detector measuremen...
متن کاملبررسی اثر ناهمگنی بافت در دوز دریافتی پرتوهای اورتو ولتاژ در ناحیه نخاع گردنی موش صحرایی
Background & Aim: The most important physical quantity of irradiation to predict and evaluate the response of biological samples to irradiation is absorbed dose. Recently, according to ICRU(International Commission on Radiation Units and Measurements) reports, determination of absorbed dose with 3.5% accuracy is recommended. Dose distribution is usually estimated in homogeneous mass by unit...
متن کاملMicro Irradiator / Animals Advanced Applications of 3D Dosimetry and 3D Printing in Radiation Therapy
Pre-clinical animal studies are used as a conventional means of translational research, as a midpoint between in-vitro cell studies and clinical implementation. However, modern small animal radiotherapy platforms are primitive in comparison with conventional linear accelerators. This work also investigates a series of 3D printed tools to expand the treatment capabilities of the X-RAD 225Cx orth...
متن کاملBiological characterization of a novel in vitro cell irradiator
To evaluate the overall robustness of a novel cellular irradiator we performed a series of well-characterized, dose-responsive assays to assess the consequences of DNA damage. We used a previously described novel irradiation system and a traditional 137Cs source to irradiate a cell line. The generation of reactive oxygen species was assessed using chloromethyl-H2DCFDA dye, the induction of DNA ...
متن کامل