AAPM Task Group 108: PET and PET/CT shielding requirements.

نویسندگان

  • Mark T Madsen
  • Jon A Anderson
  • James R Halama
  • Jeff Kleck
  • Douglas J Simpkin
  • John R Votaw
  • Richard E Wendt
  • Lawrence E Williams
  • Michael V Yester
چکیده

The shielding of positron emission tomography (PET) and PET/CT (computed tomography) facilities presents special challenges. The 0.511 MeV annihilation photons associated with positron decay are much higher energy than other diagnostic radiations. As a result, barrier shielding may be required in floors and ceilings as well as adjacent walls. Since the patient becomes the radioactive source after the radiopharmaceutical has been administered, one has to consider the entire time that the subject remains in the clinic. In this report we present methods for estimating the shielding requirements for PET and PET/CT facilities. Information about the physical properties of the most commonly used clinical PET radionuclides is summarized, although the report primarily refers to fluorine-18. Typical PET imaging protocols are reviewed and exposure rates from patients are estimated including self-attenuation by body tissues and physical decay of the radionuclide. Examples of barrier calculations are presented for controlled and noncontrolled areas. Shielding for adjacent rooms with scintillation cameras is also discussed. Tables and graphs of estimated transmission factors for lead, steel, and concrete at 0.511 MeV are also included. Meeting the regulatory limits for uncontrolled areas can be an expensive proposition. Careful planning with the equipment vendor, facility architect, and a qualified medical physicist is necessary to produce a cost effective design while maintaining radiation safety standards.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effective dose to patients and staff when using a mobile PET/SPECT system

The purpose of this study was to determine the number of weekly acquisitions permissible using a mobile PET/SPECT scanner for myocardial perfusion/viability imaging in an intensive care unit (ICU) based on the effective dose to patients and staff. The effective dose to other patients and staff in an ICU was calculated following recommendations from the American Association of Physicists in Medi...

متن کامل

Evaluation of Radiation Exposure to Staff and Environment Dose from [18F]-FDG in PET/CT and Cyclotron Center using Thermoluminescent Dosimetry

Background: PET/CT imaging using [18F]-FDG is utilized in clinical oncology for tumor detecting, staging and responding to therapy procedures. Essential consideration must be taken for radiation staff due to high gamma radiation in PET/CT and cyclotron center. The aim of this study was to assess the staff exposure regarding whole body and organ dose and to evaluate environment dose in PET/CT an...

متن کامل

Toward a standard for the evaluation of PET‐Auto‐Segmentation methods following the recommendations of AAPM task group No. 211: Requirements and implementation

PURPOSE The aim of this paper is to define the requirements and describe the design and implementation of a standard benchmark tool for evaluation and validation of PET-auto-segmentation (PET-AS) algorithms. This work follows the recommendations of Task Group 211 (TG211) appointed by the American Association of Physicists in Medicine (AAPM). METHODS The recommendations published in the AAPM T...

متن کامل

A Comparative Study of the Construction of Positron Emission Tomography/Computed Tomography Facilities in Three South African Hospitals

Introduction: Development of higher energy modalities such as positron emission tomography/computed tomography (PET/CT), has led to more complex shielding problems. This is due to several factors, such as the radiopharmaceutical relatively high-administered activity, high patient throughput, and high energies of 511 kilo-electron volt (keV) positron annihilation photon...

متن کامل

Diagnostic Accuracy of Positron Emission Mammography with 18F-fluorodeoxyglucose in Breast Cancer Tumor of Less than 20 mm in Size

Objective(s): To investigate the diagnostic accuracy of positron emission mammography (PEM) and positron emission tomography/computed tomography (PET/CT) for small breast tumors of less than 20 mm in size.Methods: The study was conducted on a total of 100 subjects (i.e., 50 patients with pathologically proven breast cancer and 50 normal cases of medical screening). The total number of tumors wa...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 33 1  شماره 

صفحات  -

تاریخ انتشار 2006