Ventolin Diskus and Inspyril Turbuhaler: An In Vitro Comparison

نویسنده

  • Mary Ann Liebert
چکیده

Dose delivery (total emitted dose, or TED) from dry powder inhalers (DPIs), pulmonary deposition, and the biological effects depend on drug formulation and device and patient characteristics. The aim of this study was to measure, in vitro, the relationship between parameters of inhalation profiles recorded from patients, the TED and fine particle mass (FPM) of Diskus and Turbuhaler inhalers. Inhalation profiles (IPs) of 25 patients, a representative sample of a wide range of 1500 IPs generated by 10 stable asthmatics, 3 16 (mild/moderate/severe) COPD patients and 15 hospitalized patients with an exacerbation asthma or COPD, were selected for each device. These 25 IPs were input IPs for the Electronic Lung (a computerdriven inhalation simulator) to determine particle size distribution from Ventolin Diskus and Inspyril Turbuhaler. The TED and FPM of Diskus and FPM of Turbuhaler were affected by the peak inspiratory flow (PIF) and not by slope of the pressure-time curve, inhaled volume and inhalation time. This flow-dependency was more marked at lower flows (PIF 40 L/min). Both the TED and FPM of Diskus were significantly higher as compared to those of the Turbuhaler [mean (SD) TED_diskus (%label claim) 83.5 (13.9) vs. TED_turbuhaler (72.5 (11.1) (p 0.004), FPM_diskus (%label claim) 36.8 (9.8) vs FPM_turbuhaler (28.7 (7.7) (p 0.05)]. The TED and FPM of Diskus and FPM of Turbuhaler were affected by PIF, the flow-dependency being greater at PIF values below 40 L/min. Lower PIFs occurred more often when using Turbuhaler than Diskus, since Turbuhaler have a higher resistivity, requires substantially higher pressure in order to generate the same flow as Diskus. TED, dose consistency and the FPM were higher for Diskus as compared to Turbuhaler. The flow dependency of TED and FPM was substantially influenced by inhalation profiles when not only profiles of the usual outpatient population were included but also the real outliers from exacerbated patients.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Ventolin Diskus and Inspyril Turbuhaler: an in vitro comparison.

Dose delivery (total emitted dose, or TED) from dry powder inhalers (DPIs), pulmonary deposition, and the biological effects depend on drug formulation and device and patient characteristics. The aim of this study was to measure, in vitro, the relationship between parameters of inhalation profiles recorded from patients, the TED and fine particle mass (FPM) of Diskus and Turbuhaler inhalers. In...

متن کامل

User error with Diskus and Turbuhaler by asthma patients and pharmacists in Jordan and Australia.

BACKGROUND Use of inhalers requires accurate completion of multiple steps to ensure effective medication delivery. OBJECTIVE To evaluate the most problematic steps in the use of Diskus and Turbuhaler for pharmacists and patients in Jordon and Australia. METHODS With standardized inhaler-technique checklists, we asked community pharmacists to demonstrate the use of Diskus and Turbuhaler. We...

متن کامل

Fine particle mass from the Diskus inhaler and Turbuhaler inhaler in children with asthma.

The study aimed to investigate dose consistency and particle distribution from the dry powder inhalers Diskus and Turbuhaler. Full profiles of inhalation pressure versus time were recorded in 18 4 yr old and 18 8 yr old asthmatic children through Diskus and Turbuhaler inhalers. These data were used in an inhalation profile simulator to assess drug delivery from both a Diskus inhaler and a Turbu...

متن کامل

Peak inspiratory flow rate and slope of the inhalation profiles in dry powder inhalers.

Aerosol delivery depends on device design and inhalation technique. In vitro device evaluations have shown that the emitted dose and fine particle mass of dry powder inhalers (DPIs) increase at high peak inspiratory flow rates (PIFR). Since the PIFR is mostly achieved after the release of the powder, slope of the pressure profile is also described as an important determinant. The aim of the pre...

متن کامل

Lung deposition and systemic availability of fluticasone Diskus and budesonide Turbuhaler in children.

Pharmacokinetic studies can be used to measure lung dose of inhaled drugs. The aim of this study was to compare the lung deposition of budesonide (BUD) inhaled from Turbuhaler (AstraZeneca, Lund, Sweden) and fluticasone propionate (FP) inhaled from Diskus (GlaxoSmithKline, London, UK) and to assess if the study design used for pharmacokinetic studies can be simplified. Plasma levels of BUD and ...