Do Catheter Side Holes Provide Better Blood Flows?
نویسندگان
چکیده
Four catheters (Ash Split Cath, Tesio, Duo-Split, and Duo--Flow; Medcomp, Harleysville, PA, U.S.A.) were tested in a temperature-controlled in vitro setup filled with 50% aqueous glycerin solution to determine hydraulic resistance at different flow rates. All these catheters have side holes; hydraulic resistance was determined with these holes open and closed. Due to extra pressure losses near the catheter tip, the pressure-flow relationship deviates from Poiseuillian theory and is generally quadratic in nature. An equivalent diameter was derived from the data. This equivalent diameter can be used to evaluate performance using a single number. Permanent catheters can easily deliver 300 mL/minute under optimal circumstances, but acute catheters are, in practice, limited to 200 mL/minute, and even somewhat less in the coaxial Duo-Flow type. Permanent catheters have larger equivalent internal diameters (1.8 vs 1.45 mm). Covering the side holes does not influence hydraulic resistance to a great degree, except in the arterial limb of acute catheters. These results indicate that, especially in acute catheters, obstruction of the side holes or fibrin sleeve/thrombus formation over the inlet holes may severely impact the available blood flow rate during dialysis. On the other hand, side holes in permanent catheters or venous limbs seem to be superfluous for performance reasons.
منابع مشابه
Fluid mechanics and clinical success of central venous catheters for dialysis--answers to simple but persisting problems.
Over 60% of patients initiating chronic hemodialysis in the United States have a chronic central venous catheter (CVC) as their first blood access device. Although it would be better if these patients started dialysis with fistulas, the CVC is used because it is a reliable and relatively safe method for obtaining blood access over a period of months. Drawing blood from a vein at 300-400 ml/minu...
متن کاملA novel catheter for simultaneous angiography of ipsilateral pulmonary veins.
We designed a novel catheter for simultaneous angiography of ipsilateral pulmonary veins (PVs), so that the catheter tip with side holes can be introduced into the inferior PVs and a second curve with side holes into the superior PVs. The simultaneous angiogram of the ipsilateral PVs, using this novel catheter, reveals more clearly the PV ostium to antrum, because simultaneous injection of cont...
متن کاملAnesthesia for endobronchial laser surgery: a modified technique.
UNLABELLED We describe a technique for endobronchial surgery with the neodynium:yttium-aluminum-garnet laser, in which an insufflation catheter with side holes placed into the contralateral mainstem bronchus is used for high-frequency positive pressure ventilation. Thirty-five patients (45 procedures) were treated during general anesthesia using a rigid bronchoscope in combination with a fibero...
متن کاملBlood pressure control with fenoldopam during excision of a pheochromocytoma.
minimize the risk of barotrauma, the following precautions are recommended. multiple distal side holes also may decrease the pressure delivered at the distal end, minimize the catheter whip, and center the catheter within the trachea during jet ventilation.3 1. The patient must be adequately preoxygenated and hyperventilated by the original endotracheal tube using 100% oxygen. 2. An exchange ca...
متن کاملCalculation of the relativistic bulk tensor and shear tensor of relativistic accretion flows in the Kerr metric.
In this paper, we calculate the relativistic bulk tensor and shear tensor of the relativistic accretion ows in the Kerr metric, overall and without any approximation. We obtain the relations of all components of the relativistic bulk and shear tensor in terms of components of four-velocity and its derivatives, Christoffel symbols and metric components in the BLF. Then, these components are deri...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Hemodialysis international. International Symposium on Home Hemodialysis
دوره 6 1 شماره
صفحات -
تاریخ انتشار 2002