Lens thickness assessment: anterior segment optical coherence tomography versus A-scan ultrasonography.

نویسندگان

  • Nikoo Hamzeh
  • Sasan Moghimi
  • Golshan Latifi
  • Massood Mohammadi
  • Nassim Khatibi
  • Shan C Lin
چکیده

AIM To assess lens thickness measurements with anterior segment-optical coherence tomography (AS-OCT) in comparison with A-scan ultrasonography (A-scan US). METHODS There were 218 adult subjects (218 eyes) aged 59.2±9.2y enrolled in this prospective cross-sectional study. Forty-three eyes had open angles and 175 eyes had narrow angles. Routine ophthalmic exam was performed and nuclear opacity was graded using the Lens Opacities Classification System III (LOCS III). Lens thickness was measured by AS-OCT (Visante OCT, Carl Zeiss Meditec, Dublin, CA, USA). The highest quality image was selected for each eye and lens thickness was calculated using ImageJ software. Lens thickness was also measured by A-scan US. RESULTS Interclass correlations showed a value of 99.7% for intra-visit measurements and 95.3% for inter-visit measurements. The mean lens thickness measured by AS-OCT was not significantly different from that of A-scan US (4.861±0.404 vs 4.866±0.351 mm, P=0.74). Lens thickness values obtained from the two instruments were highly correlated overall (Pearson correlation coefficient=0.81, P<0.001), and in all LOCS III specific subgroups except in grade 5 of nuclear opacity. Bland-Altman analysis revealed a 95% limit of agreement from -0.45 to 0.46 mm. Lens thickness difference between the two instruments became smaller as the lens thickness increased and AS-OCT yielded smaller values than A-scan US in thicker lens (β=-0.29, P<0.001). CONCLUSION AS-OCT-derived lens thickness measurement is valid and comparable to the results obtained by A-scan US. It can be used as a reliable noncontact method for measuring lens thickness in adults with or without significant cataract.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Anterior segment optical coherence tomography and retained vegetal intraocular foreign body masquerading as chronic anterior uveitis

BACKGROUND The purpose of this single case report was to report the use of anterior segment optical coherence tomography for the diagnosis and management of a retained vegetal intraocular foreign body. RESULTS A 23-year-old otherwise healthy male presented with a progressive vision loss in the right eye (RE). He reported a mild ocular trauma with a tree leaf 1 year ago followed by recurrent e...

متن کامل

Full anterior segment biometry with extended imaging range spectral domain optical coherence tomography at 1340 nm.

We demonstrate an extended-imaging-range anterior-segment optical coherence tomography (eAS-OCT) system for the biometric assessment of full AS in human eye. This newly developed eAS-OCT operating at 1340-nm wavelength band is simultaneously capable of an imaging speed of 120 kHz A-line scan rate, an axial resolution of 7.2 μm, and an extended imaging range of up to 16 mm in air. Imaging result...

متن کامل

Changes in the anterior segment after cycloplegia with a biometer using swept-source optical coherence tomography

The aim of this study was to investigate changes in the anterior segment of the eye after cycloplegia. A biometer combined with swept-source optical coherence tomography (SSOCT) was used for measurements. Patients with strabismus or amblyopia who underwent cycloplegia were included. The axial length, central corneal thickness, anterior chamber depth, and lens thickness were measured with the bi...

متن کامل

Optical Coherence Tomography Reveals New Insights into the Accommodation Mechanism

Purpose. To evaluate the movement of the anterior and posterior lens poles during naturally stimulated accommodation in children using anterior segment optical coherence tomography (OCT). Methods. This is a prospective, observational, noncomparative case series including 18 eyes of nine children. Analysis of the anterior segment in the accommodated and unaccommodated state (with cycloplegia) wa...

متن کامل

Optical coherence tomography assessment of angle anatomy changes after cataract surgery.

PURPOSE To evaluate changes in anterior chamber (AC) morphology induced by cataract extraction using anterior segment optical coherence tomography (OCT). DESIGN Prospective comparative observational case series. METHODS Thirty-two eyes of 32 patients underwent OCT imaging of the angle before and after cataract surgery. Anterior chamber depth (ACD), angle opening distance at 500 microm (AOD(...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • International journal of ophthalmology

دوره 8 6  شماره 

صفحات  -

تاریخ انتشار 2015