Measuring peak expiratory flow in general practice: comparison of mini Wright peak flow meter and turbine spirometer.

نویسندگان

  • K P Jones
  • M A Mullee
چکیده

OBJECTIVE To compare measurements of the peak expiratory flow rate taken by the mini Wright peak flow meter and the turbine spirometer. DESIGN Pragmatic study with randomised order of use of recording instruments. Phase 1 compared a peak expiratory flow type expiration recorded by the mini Wright peak flow meter with an expiration to forced vital capacity recorded by the turbine spirometer. Phase 2 compared peak expiratory flow type expirations recorded by both meters. Reproducibility was assessed separately. SETTING Routine surgeries at Aldermoor Health Centre, Southampton. SUBJECTS 212 Patients aged 4 to 78 presenting with asthma or obstructive airways disease. Each patient contributed only once to each phase (105 in phase 1, 107 in phase 2), but some entered both phases on separate occasions. Reproducibility was tested on a further 31 patients. MAIN OUTCOME MEASURE 95% Limits of agreement between measurements on the two meters. RESULTS 208 (98%) Of the readings taken by the mini Wright meter were higher than the corresponding readings taken by the turbine spirometer, but the 95% limits of agreement (mean difference (2 SD] were wide (1 to 173 l/min). Differences due to errors in reproducibility were not sufficient to predict this level of disagreement. Analysis by age, sex, order of use, and the type of expiration did not detect any significant differences. CONCLUSIONS The two methods of measuring peak expiratory flow rate were not comparable. The mini Wright meter is likely to remain the preferred instrument in general practice.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Peak Expiratory Flow Rate in Healthy Children from Tehran

Background: Measuring peak expiratory flow rate has been suggested as an important tool in asthma management by all international guidelines.  Objective: Studies on various populations have shown considerable variations in normal values of peak expiratory flow rate. Since reference standards for Iranian children are not available, the present study was performed to establish these standards. Me...

متن کامل

Evaluation of the turbine pocket spirometer.

A compact electronic spirometer, the turbine pocket spirometer, which measures the FEV1, forced vital capacity (FVC), and peak expiratory flow (PEF) in a single expiration, was compared with the Vitalograph and the Wright peak flow meter in 99 subjects (FEV1 range 0.40-5.50 litres; FVC 0.58-6.48 l; PEF 40-650 l min-1). The mean differences between the machines were small--0.05 l for FEV1, 0.05 ...

متن کامل

Peak expiratory flow at altitude.

The mini Wright peak flow meter is a useful, portable instrument for field studies but being sensitive to air density will under-read at altitude. True peak expiratory flow will increase at altitude, however, because of the decreased air density, given that dynamic resistance is unchanged. The effect of simulated altitude on peak expiratory flow (PEF) was determined in six subjects with both th...

متن کامل

Assessment of accuracy and applicability of a portable electronic diary card spirometer for asthma treatment.

A pocked-sized turbine flowmeter and spirometer device, integrated with an electronic diary card (EDC-spirometer, Micro Medical, U.K.), was tested with a mechanical calibrator, in an outpatient clinic and in the home situation. A screen pneumotachometer was used as flow and volume reference. Ten devices were tested; interdevice variability was small with a mean variation coefficient of 1.1% for...

متن کامل

Effects of equipment and technique on peak flow measurements

BACKGROUND Different lung function equipment and different respiratory manoeuvres may produce different Peak Expiratory Flow (PEF) results. Although the PEF is the most common lung function test, there have been few studies of these effects and no previous study has evaluated both factors in a single group of patients. METHODS We studied 36 subjects (PEF range 80-570 l/min). All patients reco...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • BMJ

دوره 300 6740  شماره 

صفحات  -

تاریخ انتشار 1990