Combination of converging collimators for high-sensitivity brain SPECT.
نویسندگان
چکیده
UNLABELLED The objective of this study, which is related to human brain SPECT, was to increase the sensitivity of a triple-camera SPECT system and reduce statistical noise in reconstructed images using a combination of converging collimators. The reason for combining collimators is to ensure both high sensitivity and sufficient sampling without trading off spatial resolution. METHODS A high-sensitivity half-cone-beam (HCB) collimator, designed specifically for brain imaging, was combined with other collimators and compared with conventional parallel-beam and fanbeam circular orbit acquisitions. For comparison, previously studied HCB collimation with a circle-and-helix data acquisition trajectory was also included in this study. Simulations of the Hoffman 3-dimensional brain phantom were performed to calculate the efficiencies of collimators and their combinations and to quantitatively evaluate reconstruction bias, statistical noise, and signal-to-noise ratios in the reconstructed images. Experimental brain phantom data were also acquired and compared for different acquisition types. Finally, a patient brain scan was obtained with a combination of HCB and fanbeam collimators and compared with a triple-fanbeam circular orbit acquisition. RESULTS A combination of 2 HCB collimators and 1 fanbeam collimator, compared with a triple-fanbeam collimator, can increase the photon detection efficiency by 27% and by more than a factor of 2, compared with triple-parallel-hole collimation, with equal spatial resolution measured on the axis of rotation. Quantitative analysis of reconstruction bias and visual analysis of the images showed no signs of sampling artifacts. Reconstructed images in the simulations, experimental brain phantom, and patient brain scans showed improved quality with this collimator combination due to increased sensitivity and reduced noise. Lesion visibility was also improved, as confirmed by signal-to-noise ratios. Alternatively, triple-HCB circle-and-helix acquisition has also shown competitive results, with a slight disadvantage in axial sampling and implementation procedure. CONCLUSION Combined HCB and fanbeam collimation is a promising approach for high-sensitivity brain SPECT.
منابع مشابه
Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT Images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators
Objective(s): The purpose of this study was to examine the optimal reconstruction parameters for brain dopamine transporter SPECT images obtained with a fan beam collimator and compare the results with those obtained by using parallel-hole collimators.Methods: Data acquisition was performed using two SPECT/CT devices, namely a Symbia T6 and an Infinia Hawkeye 4 (device A and B) equipped with fa...
متن کاملOptimization of an ultra-high-resolution rectangular pixelated parallel-hole collimator with a CZT pixelated semiconductor detector for HiRe-SPECT system
Introduction: In nuclear medicine, the use of a pixelated semiconductor detector such as CZT is an of growing interest for introducing new devices. Especially, the spatial resolution can be improved by using a pixelated parallel-hole collimator with equal holes and pixel sizes based on the pixelated detector. The purpose of this study was to compare the effect of pixelated and ...
متن کاملTomographic image quality of rotating slat versus parallel hole-collimated SPECT.
Parallel and converging hole collimators are most frequently used in nuclear medicine. Less common is the use of rotating slat collimators for single photon emission computed tomography (SPECT). The higher photon collection efficiency, inherent to the geometry of rotating slat collimators, results in much lower noise in the data. However, plane integrals contain spatial information in only one ...
متن کاملOptimization of a SPECT system for imaging of 90Y in liver using Monte Carlo method
Introduction: Acquiring a high quality image has assigned an important concern for obtaining accurate diagnosis in nuclear medicine. Detector and collimator are critical component of Single Photon Emission Computed Tomography (SPECT) imaging system for giving accurate information from exact pattern of radionuclide distribution in the target organ. The images are strongly affect...
متن کاملSPECT using a specially designed cone beam collimator.
A specially designed high resolution converging collimator having a focal length of 50 cm has been evaluated for cone beam single photon emission computed tomography (SPECT). The focal region was investigated by imaging a point source placed at the expected focal point and along the central ray of the collimator in front of and behind the focal point. Technetium-99m point source sensitivities m...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of nuclear medicine : official publication, Society of Nuclear Medicine
دوره 50 9 شماره
صفحات -
تاریخ انتشار 2009