Correction for vascular artifacts in cerebral blood flow values measured by using arterial spin tagging techniques.

نویسندگان

  • F Q Ye
  • V S Mattay
  • P Jezzard
  • J A Frank
  • D R Weinberger
  • A C McLaughlin
چکیده

"Vascular" artifacts can have substantial effects on human cerebral blood flow values calculated by using arterial spin tagging approaches. One vascular artifact arises from the contribution of "tagged" arterial water spins to the observed change in brain water MR signal. This artifact can be reduced if large bipolar gradients are used to "crush" the MR signal from moving arterial water spins. A second vascular artifact arises from relaxation of "tagged" arterial blood during transit from the tagging plane to the capillary exchange site in the imaging slice. This artifact can be corrected if the arterial transit times are measured by using "dynamic" spin tagging approaches. The mean transit time from the tagging plane to capillary exchange sites in a gray matter region of interest was calculated to be approximately 0.94 s. Cerebral blood flow values calculated for seven normal volunteers agree reasonably well with values calculated by using radioactive tracer approaches.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

H(2)(15)O PET validation of steady-state arterial spin tagging cerebral blood flow measurements in humans.

Steady-state arterial spin tagging approaches can provide quantitative images of CBF, but have not been validated in humans. The work presented here compared CBF values measured using steady-state arterial spin tagging with CBF values measured in the same group of human subjects using the H(2)(15)O IV bolus PET method. Blood flow values determined by H(2)(15)O PET were corrected for the known e...

متن کامل

Validation of diffuse correlation spectroscopy measurements of rodent cerebral blood flow with simultaneous arterial spin labeling MRI; towards MRI-optical continuous cerebral metabolic monitoring

Cerebral blood flow (CBF) during stepped hypercapnia was measured simultaneously in the rat brain using near-infrared diffuse correlation spectroscopy (DCS) and arterial spin labeling MRI (ASL). DCS and ASL CBF values agree very well, with high correlation (R=0.86, p< 10(-9)), even when physiological instability perturbed the vascular response. A partial volume effect was evident in the smaller...

متن کامل

Arterial transit time imaging with flow encoding arterial spin tagging (FEAST).

Arterial spin labeling (ASL) perfusion imaging provides direct and absolute measurement of cerebral blood flow (CBF). Arterial transit time is a related physiological parameter reflecting the duration for the labeled spins to reach the brain region of interest. Most of the existing ASL approaches to assess arterial transit time rely on multiple measurements at various postlabeling delay times, ...

متن کامل

Characterizing the origin of the arterial spin labelling signal in MRI using a multiecho acquisition approach.

Arterial spin labelling (ASL) can noninvasively isolate the MR signal from arterial blood water that has flowed into the brain. In gray matter, the labelled bolus is dispersed within three main compartments during image acquisition: the intravascular compartment; intracellular tissue space; and the extracellular tissue space. Changes in the relative volumes of the extracellular and intracellula...

متن کامل

Comparison of arterial transit times estimated using FEAST and LL-FAIR

Introduction Arterial transit time is a critical parameter for calculating cerebral blood flow (CBF) using arterial spin labeling (ASL). Arterial transit time is also a novel physiological parameter that is uniquely measured by ASL and may have independent clinical significance for detecting CBF that is maintained through collateral flow sources. Most ASL experiments acquire data at a single de...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Magnetic resonance in medicine

دوره 37 2  شماره 

صفحات  -

تاریخ انتشار 1997